ASX-1400 Autosampler Operator’s Manual
Using the Autosampler
5–6
2)
Turn the ASX-1400 Autosampler power switch on.
The green LED indicator along the autosampler X-Axis lights up
when the power is on.
3)
Adjust the pump speed control knob until the desired rinse
solution flow rate is achieved.
4)
Purge air from the rinse system by placing the rinse solution
uptake tubing in the rinse solution source and running the
rinse solution through the rinse station.
Ensure there are no air bubbles visible in the rinse uptake tubing
before you run samples with the ASX-1400 Autosampler.
Note:
If you are flushing the rinse system during initial startup, first use a 2%
nitric acid solution as the rinse agent. Flush the rinse system a second
time using deionized water as the rinse agent.
5)
Access the host computer’s software and activate the
autosampler program.
The ASX-1400 Autosampler runs until it reaches the end of the
sampling sequence.
Содержание ASX-1400
Страница 1: ...ASX 1400 Autosampler Operator s Manual ...
Страница 12: ...Contents ...
Страница 17: ...Preface ...
Страница 23: ...1 Introduction ...
Страница 31: ...2 Preparing for Installation ...
Страница 35: ...3 Installing the Autosampler ...
Страница 39: ...ASX 1400 Autosampler Operator s Manual Installing the Autosampler 3 5 Figure 3 2 Mounting of Z Drive Assembly ...
Страница 43: ...ASX 1400 Autosampler Operator s Manual Installing the Autosampler 3 9 Figure 3 2 Pumped Drain Arrangement ...
Страница 52: ...4 Verifying Installation ...
Страница 60: ...5 Using the Autosampler ...
Страница 68: ...6 1 6 Maintaining the Autosampler ...
Страница 76: ...7 Troubleshooting the Autosampler ...
Страница 81: ...8 Appendix A ...