10 APPENDIX A- Technical references
Clause
6.8.3.201 Technical Description - Tab 204 Guidance and manufacturer's declaration
- electromagnetic immunity - for equipment and systems that are not life-supporting
TABLE: Guidance and manufacturer's declaration - electromagnetic immunity
The device
NewTom 5G
is intended for use in the electromagnetic environment specified below. The
customer or the user of the device
NewTom 5G
should assure that is used in such an environment.
Immunity test
IEC 60601 Test Level
Compliance Level
Electromagnetic environment - guidance
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3 Vrms
150 kHz to 80 MHz
3 V/m
80 MHz to 2,5 GHz
3 Vrms
3 V/m
Portable and mobile RF communications
equipment should be used no closer to any part
of the
NewTom 5G
, including cables, than the
recommended separation distance calculated
from the equation applicable to the frequency of
the transmitter.
Recommended separation distance:
d
= 1.2
Ö
P
d
= 1.2
Ö
P
da 80 MHz a 800 MHz
d
= 2.3
Ö
P
da 800 MHz a 2.5 GHz
where
P
is the maximum output power rating of
the transmitter in watts (W) according to the
transmitter manufacturer and
d
is the
recommended separation distance in meters(m).
Fields strengths from fixed RF transmitters, as
determined by an electromagnetic site survey*,
should be less than the compliance level in each
frequency range**.
Interference may occur in the vicinity of
equipment marked with the following symbol:
NOTE:
1)
At 80 MHz and 800MHz, the separation distance for the higher frequency range applies
2)
These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from
structures, objects and people.
*
Field strengths from fixed transmitters, such as base stations for radio (celular/cordless), telephones and land mobile radios, amateur
radio, AM and FM radio broadcast, and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic
environment due to fixed RF transmitters, an electromagnetic site survey should be performed. If the measured field strength in the
location in which the
NewTom 5G
is used
exceeds the RF compliance level above, the
NewTom 5G
should be observed to verify
normal operation. If abnormal performance is observed, additional measures may be necessary, such as re-orienting or relocating
NewTom 5G.
**Over the frequency range 150 kHz to 80 MHz, field strhgths should be less than 3 V/m.
10-26
NewTom 5G
– User Manual
Содержание NewTom 5G
Страница 1: ...NewTom 5G User Manual 0051 97050196 Rev 5 28 04 2017 EN...
Страница 80: ...10 APPENDIX A Technical references NewTom 5G User Manual 10 7...
Страница 81: ...10 APPENDIX A Technical references 10 8 NewTom 5G User Manual...
Страница 82: ...10 APPENDIX A Technical references NewTom 5G User Manual 10 9...
Страница 83: ...10 APPENDIX A Technical references 10 10 NewTom 5G User Manual...