10 APPENDIX A- Technical references
CTDIvol table
Values expressed in mGy. Tolerance is
30%.
Values calculated according to IEC 60601-2-44: 2009 Par 201.3.211 and 201.3.212 c)
FOV
Eco Scan
Standard Dose
Regular Scan
Standard Dose
Regular Scan
Boosted Dose
Enhanced Scan
Standard Dose
Enhanced Scan
Boosted Dose
[15x22e]
4,6
7,0
10,9
-
-
35
Loading factors: automaticamente impostati dal software
Eco Scan Std Dose: 110kV 32,85mAs - Reg. Scan Std Dose: 110kV 48,84mAs - Reg. Scan Boosted Dose: 110kV 97,63mAs