3 Safety and maintenance of the device
Responsable
Part
Activity
Frequency
Routine testing
User
Global system
QA Phantom check
Weekly
Local Support
Service
Error log
Check
12 months
Every external part
Damage check
12 months
Emergency buttons
Emergency test
12 months
Electrical functioning
Check
12 months
Mechanical functioning
Check
12 months
Other test depending of local regulations
Radioprotection
expert or other
qualified person
depending on
local regulations
Global system
Radiological test
conforming to the local
regulation concerning X
-ray
medical electrical
equipment.
This tests in not in charge
by user or local support
service but may be
established by local
regulations.
Radiological test
conforming to the
local standard
NewTom 5G
– User Manual
3-5