10 APPENDIX A- Technical references
CTDI
free air
table declared for operative modalities of the device
Values expressed in mGy. Tolerance is
30%. Values calculated according to IEC 60601-2-44: 2009 Par 203.109.2
FOV
Eco Scan
Standard Dose
Regular Scan
Standard Dose
Regular Scan
Boosted Dose
Enhanced Scan
Standard Dose
Enhanced Scan
Boosted Dose
[18x16]
4,9
6,7
13,7
11,2
18,1
[15x22e]
7,0
10,4
17,3
-
-
[15x12]
5,2
7,0
14,6
11,9
19,6
[12x8]
7,3
9,6
15,5
16,3
20,7
[8x8]
7,7
10,6
16,6
17,3
22,2
[15x5] HiRes
18,9
28,2
-
38,3
-
[12x8] HiRes
16,6
25,0
-
33,5
-
[8x8] HiRes
16,8
25,1
-
34,0
-
[6x6] HiRes
19,7
24,4
-
39,9
-
Date: 2013-07-12
46
Loading factors: automatically set by software
Eco Scan Std Dose: 110kV 17.76mAs - Reg. Scan Std Dose: 110kV 26.35mAs - Reg. Scan Boosted Dose: 110kV 67.20mAs
- Enh. Scan Std Dose: 110kV 35.52mAs - Enh. Scan Boosted Dose: 110kV 87.64mAs
47
Loading factors: automaticamente impostati dal software
Eco Scan Std Dose: 110kV 32,88mAs - Reg. Scan Std Dose: 110kV 48,88mAs - Reg. Scan Boosted Dose: 110kV 97,63mAs
48
Loading factors: automatically set by software
Eco Scan Std Dose: 110kV 19.07mAs - Reg. Scan Std Dose: 110kV 28.37mAs - Reg. Scan Boosted Dose: 110kV 71.89mAs
- Enh. Scan Std Dose: 110kV 38.14mAs - Enh. Scan Boosted Dose: 110kV 97.61mAs
49
Loading factors: automatically set by software
Eco Scan Std Dose: 110kV 23.14mAs - Reg. Scan Std Dose: 110kV 36.00mAs - Reg. Scan Boosted Dose: 110kV 64.21mAs
- Enh. Scan Std Dose: 110kV 46.28mAs - Enh. Scan Boosted Dose: 110kV 85.61mAs
50
Loading factors: automatically set by software
Eco Scan Std Dose: 110kV 25.30mAs - Reg. Scan Std Dose: 110kV 39.55mAs - Reg. Scan Boosted Dose: 110kV 71.25mAs
- Enh. Scan Std Dose: 110kV 50.60mAs - Enh. Scan Boosted Dose: 110kV 93.27mAs
51
Loading factors: automatically set by software
Eco Scan Std Dose: 110kV 72.60mAs - Reg. Scan Std Dose: 110kV 108.90mAs - Enh. Scan Std Dose: 110kV 147.01mAs
52
Loading factors: automatically set by software
Eco Scan Std Dose: 110kV 72.60mAs - Reg. Scan Std Dose: 110kV 108.90mAs - Enh. Scan Std Dose: 110kV 147.01mAs
53
Loading factors: automatically set by software
Eco Scan Std Dose: 110kV 72.60mAs - Reg. Scan Std Dose: 110kV 108.90mAs - Enh. Scan Std Dose: 110kV 147.01mAs
54
Loading factors: automatically set by software
Eco Scan Std Dose: 110kV 72.60mAs - Reg. Scan Std Dose: 110kV 89.84mAs - Enh. Scan Std Dose: 110kV 147.01mAs
NewTom 5G
– User Manual
10-19
Содержание NewTom 5G
Страница 1: ...NewTom 5G User Manual 0051 97050196 Rev 5 28 04 2017 EN...
Страница 80: ...10 APPENDIX A Technical references NewTom 5G User Manual 10 7...
Страница 81: ...10 APPENDIX A Technical references 10 8 NewTom 5G User Manual...
Страница 82: ...10 APPENDIX A Technical references NewTom 5G User Manual 10 9...
Страница 83: ...10 APPENDIX A Technical references 10 10 NewTom 5G User Manual...