10 APPENDIX A- Technical references
All components, accessories, spare parts must be approved and supplied by CEFLA s.c.
Particularly, the connection cables must be ONLY the ones described under Par. 4.4.1 - “Cables“.
WARNING:
Use of accessories, transducers, and cables other than those
specified may result in degraded electromagnetic
compatibility performance of this device!
WARNING:
NewTom 5G should not be stacked directly on top of other
equipment, and other equipment should not be stacked on
top of NewTom 5G. If stacking is necessary, observe
NewTom 5G to verify normal operation in the stacked
configuration in which it will be used!
10.11.1
Essential performance
In case of a scan stopped by a temporary or permanent malfunctioning, the operator will have the
opportunity to store the data acquired until the failure appeared.
The quality of the reconstructed images will vary depending on the size of acquired data but it will be
probably lower compared to the standard performance of the equipment.
10-28