The guarantee does not cover parts subject to wear
and repairs or replacements effected by personnel
that is not authorized by C.B.M.
During
the
guarantee
period
the
possible
replacement of an autoclave will be decided by
C.B.M. only in case it is impossible to effect a repair
such to assure the correct running of the machine.
Data,descriptions and illustrations are purely
indicative.
C.B.M. S.r.l. reserves the right to carry out
ameliorative modifications to its own products without
any previous notice.
Disposal Of Electric And Electronic
Refuse
Waste from Electrical and Electronic
Equipment - Directive 2002/96/EC
(WEEE)
The symbol (shown above) that appears on this
product means that it should not be disposed of
together with normal household refuse but should be
treated separately.
Electric and electronic equipment may contain
substances that are harmful to the environment and
to the health of humans and should therefore be
disposed of at special waste disposal sites.
There are harsh legal penalties in force against
persons found disposing of such equipment in an
irresponsible manner.
When the equipment is no longer of any use, it may
taken back to the outlet from which you intend to
purchase a replacement.
STERILIZATION TIMES
Temp.(
°
C)
Time-
Unpacked instruments-
134
°
C
5min.
Packed instruments/Glass -
134
°
C
20min.
Waste in stainless container
134
°
C
30min.
Porous material/ rubber not packed
121
°
C
20min.
Porous material/ rubber packed
121
°
C
30min.
Emergency rapid cycle for unpacked
instruments
134
°
C
3min.
DRYING
Available drying times
0 - 10 - 15 - 20 - 25 - 30min.
ATTENTION: In order to have an optimal working of the autoclave follow the instructions
below:
1) In case of sterilization of surgical instruments or metallic material do not exceed a
quantity of 5 Kg.
2) In case of sterilization of textiles it is a good rule not do obstruct the bottom grate holes
completely and not fill the chamber beyond 80% of the available volume.