C.B.M. S.r.l. Medical Equipment - Via Castello 10, 26038 Torre d
é
Picenardi (CR) - Italy
Rev. 7
–
24/07/2006
IGEA
Pag. 13 di 45
If the drying has been selected the writing DRYING
will appear; at this moment half-close the door 2cm.,
leaving it placed near the closing device and possibly
stopped by a turn of screw.
At the end of the drying phase the writing OK
FINISHED CYCLE will appear on the display; with a
sound signal, then open the door and take out the
sterilized material.
UTILITY FUNCTIONS
When the autoclave is in the waiting phase (never
during the execution of a sterilizing cycle) it is
possible to enter the utility functions through the
UTILITY
key. Pressing this key all the headings listed
here below are visualized on the display:
MAINTENANCE:
It indicates maintenance operations to effect and
their relating terms.
TEST LEDS:
It effects a functional inspection of the leds.
TEMPERATURE IN THE CHAMBER:
It displays the actual temperature inside the
chamber.
LANGUAGE SELECTION:
It makes possible to choose the language of
messages visualized on the display among the
following languages: ITALIAN, ENGLISH, FRENCH,
GERMAN, *SPANISH
(*) if existing in the version.
DATE AND HOUR:
It makes possible to modify the dater and the clock.
Instructions for use:
After selecting the desired heading with the
UTILITY
key operate in the following way:
MAINTENANCE
:
If you wish to screen maintenance messages press
ENTER.
The 1st message will appear on the display;
in order to screen the following messages press
again the
UTILITY key
. If you wish to end the
reading before its time press the key
SHIFT
and the
RESET
key at the same time.
TEST LEDS:
In order to check the actual working of the leds press
the
ENTER
key. All the working leds will light up for 5
seconds.
TEMPERATURE IN THE CHAMBER:
In order to verify the temperature inside the chamber
press
ENTER
. Press again the same push button to go out
of the function.
LANGUAGE SELECTION:
In case you wish to modify the language press
ENTER
. Choose the desired language by the
UTILITY
key
.
To confirm your choice press
ENTER
.
DATE AND HOUR:
In order to modify date and hour press the
ENTER
key
twice
, at this moment the
CYCLE
and
DRYING
keys will work respectively from
plus
(+)
and
minus
(-
)
.
Confirm each single modified value through the
ENTER
key.
RESET AND COUNTER KEYS
If they are pressed together with the
SHIFT
key
they
take on the following functions:
RESET
1. It stops a cycle in progress, whenever any reason
makes it necessary.
2. It set alarms to zero resetting the equipment.
COUNTER
1. It displays the total number of cycles executed by
the sterilizer, at any moment
CLEANING
In order to do a good cleaning use neutral detergents
only and rinse with plenty of water.
Gaskets and all plastic parts can be cleaned with
some cotton soaked with alcohol.
Do not use sponges or abrasive materials that can
damage the surface.
During the cleaning avoid the use of detergents,
disinfectants or other liquids containing chlorides or
trichloro-ethylene because they attack the stainless
steel; in case of contact an immediate rinsing is
recommended with the replacement of the water in
the reservoir.
ROUTINE MAINTENANCE
Œ
Every 20 cycles or after two working weeks clean the
water drain filter.
ATTENTION:
Before unscrewing the filter, be sure you
have removed any dirt in the chamber
This filter is placed inside the sterilization chamber,
under the bottom grate in the middle of the
resistance and can be simply removed screwing it
out. It can be cleaned with compressed air or with
running water. It is clean when , looking into the light,
you can see its wire gauze completely free.
ATTENTION You must effect this operation every
time that the manometer indicates a residual
pressure in the chamber at the end of the
“
steam
release
”
phase, when the drying does not produce a
good result or it happens that when you open the
door some water is coming out of the chamber.
•
Every 50 cycles, or after each working month,
change the reservoir water.
In order to execute this operation:
By opening the autoclave door the proper tube can
be connected with the clutch for unloading the water
tank.
Please remember that the other end of the tube shall
be first put into a vessel to collect unloaded water.