16
CERTIFICATION OF MEDICAL EQUIPMENT ACCORDING TO DIRECTIVE 93/42/CEE
Further to the accreditation for applying CE marking to those of our appliances that are classified
as medical equipment:
ASPI-JET Models 6-7-8-9
γ
servicing authorized engineers shall use
only original
CATTANI spare parts when repairing the above
appliances.
This type of intervention (replacement of parts which have lost their mechanical/electric features in
the course of time) is considered as a corrective maintenance intervention on a breakdown with the
purpose to bring the equipment back to its initial safety condition, therefore according to the Directive
93/42 the equipment must be tested again after the intervention.
The tests to be performed are described in the standard IEC EN 62353 “Electromedical equipment”
- Periodic inspections and tests to be performed after interventions on electromedical equipment –
and it is also advisable to perform them as described in the CEI EN 60601-1 (CEI 62-5) standards.
Moreover, with reference to the components listed here below, whose lot and supplier must be easily
traced, engineers shall refer to the following table:
Components
Code
MOTOR UNI-JET 75
020354
020348
020349
020353
110 V
60 Hz
230 V
50 Hz
240 V
50 Hz
220 V
60 Hz
PRINTED BOARD
180921
180923
180930
180931
180940
180941
180943
AC 15 CIRCUIT -230 V
AC 15 CIRCUIT -110 V
Cpump -220 V
Cpump -240 V
AC 20 CIRCUIT -220 V
AC 20 CIRCUIT -240 V
AC 20 CIRCUIT -110 V
DOOR MICRO SWITCH
183102
ASS. CABLE W/MICRO
180810
While submitting an order for the above components to the sales department of CATTANI S.p.A,
they shall also indicate
SERIAL NUMBER
of the concerned appliance, committing to install the components to that appliance and not to others.