INSTALLATION
V
ENTILATION
A
SSESSMENT
R
EQUIRED
The ventilation system for the hazardous location the oven will be installed in must prevent buildups of
potentially explosive gas or vapor mixtures in the air. Conditions under which this could occur include:
•
Excessive amounts of solvents or other hazardous materials evaporating from samples
while being loaded into the oven chamber.
•
Residual hazardous materials outgassing from treated product or samples while being
unloaded from the oven after processing.
•
Explosive concentrations arising from spilled hazardous materials.
•
Contents evaporating from hazardous materials containers accidentally left open.
Determining Sufficient Ventilation
This determination must be
based on an evaluation of your site
with its airflow and HVAC
characteristics, the specifics of your baking application and the hazardous materials involved. The
HVAC or other climate control system should meet all applicable hazardous location requirements
for the safety jurisdiction you are located in.
General considerations:
•
Consult with your fire or other safety jurisdiction officials to verify the applicable ventilation
requirements and the local interpretations of those requirements.
•
The airflow or active ventilation provided at the front of the oven must be sufficient to
prevent the buildup of explosive gas or vapor concentration from samples, products, or
solvents being handled by the oven operators and other process technicians.
Neutral Gas Ventilation
•
If you are using gas nitrogen (GN
RR
2
RR
) for purging and pressurizing the oven safety
enclosures or for backfilling, ensure there is sufficient ventilation in the workspace area to
prevent asphyxiation or hypoxia conditions.
•
The ventilation must be sufficient to protect employees in the event of a leak in the GN
RR
2
supply system.
•
The workspace ventilation system must provide a minimum of
6 air changes per
minute.
Содержание CVO-5-EX Series
Страница 1: ...Flammable Solvent Ovens Installation Operation Manual CVO 5 EX Oven Series...
Страница 6: ...6 P a g e cascadesciences com Tel 503 847 9047 Special Conditions Of Use 131...
Страница 7: ...7 P a g e cascadesciences com Tel 503 847 9047 TABLE OF CONTENTS...
Страница 11: ...11 P a g e cascadesciences com Tel 503 847 9047 SAFETY CERTS AND STATEMENTS...
Страница 17: ...17 P a g e cascadesciences com Tel 503 847 9047 SPECIFICATIONS...
Страница 22: ...22 P a g e cascadesciences com Tel 503 847 9047...
Страница 24: ...24 P a g e cascadesciences com Tel 503 847 9047 These groups include propane and n butane...
Страница 29: ...29 P a g e cascadesciences com Tel 503 847 9047 Continued next page...
Страница 37: ...37 P a g e cascadesciences com Tel 503 847 9047 INTRODUCTION...
Страница 39: ...39 P a g e cascadesciences com Tel 503 847 9047...
Страница 44: ...44 P a g e cascadesciences com Tel 503 847 9047 RECEIVING...
Страница 51: ...51 P a g e cascadesciences com Tel 503 847 9047 Continued next page...
Страница 54: ...54 P a g e cascadesciences com Tel 503 847 9047 Continued next page...
Страница 61: ...61 P a g e cascadesciences com Tel 503 847 9047 End of Procedure...
Страница 65: ...65 P a g e cascadesciences com Tel 503 847 9047 INSTALLATION...
Страница 67: ...67 P a g e cascadesciences com Tel 503 847 9047 SYMBOLS...
Страница 69: ...69 P a g e cascadesciences com Tel 503 847 9047 Continued next page...
Страница 72: ...72 P a g e cascadesciences com Tel 503 847 9047 Continued next page...
Страница 95: ...95 P a g e cascadesciences com Tel 503 847 9047 End of Procedure...
Страница 102: ...102 P a g e cascadesciences com Tel 503 847 9047 Note See the next page for how to begin heating the oven chamber...
Страница 104: ...104 P a g e cascadesciences com Tel 503 847 9047 End of Procedure...
Страница 111: ...111 P a g e cascadesciences com Tel 503 847 9047 Tap the red timer Stop button to clear the Timer Finished message...
Страница 116: ...116 P a g e cascadesciences com Tel 503 847 9047 End of Procedure...
Страница 123: ...123 P a g e cascadesciences com Tel 503 847 9047 Continued next page...
Страница 125: ...125 P a g e cascadesciences com Tel 503 847 9047 Continued next page...
Страница 127: ...127 P a g e cascadesciences com Tel 503 847 9047 End of Procedure...
Страница 128: ...128 P a g e cascadesciences com Tel 503 847 9047 MAINTENANCE...
Страница 130: ...130 P a g e cascadesciences com Tel 503 847 9047 REPLACEMENT PARTS...
Страница 132: ......