71
|
P a g e
cascadesciences.com
|
Tel. 503 847-9047
CONTROLS
Note:
Backfilling is the process of restoring atmosphere to the vacuumed down oven chamber. This
is done by flowing gas into the chamber from a safe gas source connected to one of the gas-
in ports on the back of the oven.
Vacuum and Backfilling Gas Supply Valves
Vacuum Valve Control
The large valve handle labeled Vacuum adjusts the level of vacuum
draw applied to the oven chamber through the 1/4 Inch (6.35 mm)
Vacuum Port on the back of the oven.
•
When
open
, this valve allows a vacuum pump connected to
the Vacuum Port to pump down the oven chamber.
•
In the
closed
position, the valve closes and seals the port.
Vent (Bottom)
Turning the Vent handle to the open position flows gas from the Purge and Pressurize supply to
flow into the oven chamber.
•
When the Vent valve handle is in the
closed
position, the oven chamber is isolated
from the Purge and Pressurize gas supply.
•
The Vent valve should be kept in the closed position when not backfilling.
Backfill (Top)
Warning
: Do not open this valve unless the Backfill gas-in port on the back of the oven is connected
to a safe gas supply. The port must remain capped when not connected.
Avertissement:
N'ouvrez pas cette vanne à moins que l'orifice d'entrée de gaz de remplissage à
l'arrière du four ne soit connecté à une alimentation en gaz sûre. Le port doit rester fermé lorsqu'il
n'est pas connecté.
•
This valve opens and closes the
Backfill
port on the back of the oven.
•
This is intended for an optional gas supply used to backfill the oven chamber.
•
Before opening this valve, always verify
the Backfill port is connected to a safe
gas supply.
•
The Backfill port must be capped when not connected to a supply to prevent potentially
explosive gases or vapors from reaching the oven chamber.
Open
Vacuum and Vent Valves
Open
Содержание CVO-5-EX Series
Страница 1: ...Flammable Solvent Ovens Installation Operation Manual CVO 5 EX Oven Series...
Страница 6: ...6 P a g e cascadesciences com Tel 503 847 9047 Special Conditions Of Use 131...
Страница 7: ...7 P a g e cascadesciences com Tel 503 847 9047 TABLE OF CONTENTS...
Страница 11: ...11 P a g e cascadesciences com Tel 503 847 9047 SAFETY CERTS AND STATEMENTS...
Страница 17: ...17 P a g e cascadesciences com Tel 503 847 9047 SPECIFICATIONS...
Страница 22: ...22 P a g e cascadesciences com Tel 503 847 9047...
Страница 24: ...24 P a g e cascadesciences com Tel 503 847 9047 These groups include propane and n butane...
Страница 29: ...29 P a g e cascadesciences com Tel 503 847 9047 Continued next page...
Страница 37: ...37 P a g e cascadesciences com Tel 503 847 9047 INTRODUCTION...
Страница 39: ...39 P a g e cascadesciences com Tel 503 847 9047...
Страница 44: ...44 P a g e cascadesciences com Tel 503 847 9047 RECEIVING...
Страница 51: ...51 P a g e cascadesciences com Tel 503 847 9047 Continued next page...
Страница 54: ...54 P a g e cascadesciences com Tel 503 847 9047 Continued next page...
Страница 61: ...61 P a g e cascadesciences com Tel 503 847 9047 End of Procedure...
Страница 65: ...65 P a g e cascadesciences com Tel 503 847 9047 INSTALLATION...
Страница 67: ...67 P a g e cascadesciences com Tel 503 847 9047 SYMBOLS...
Страница 69: ...69 P a g e cascadesciences com Tel 503 847 9047 Continued next page...
Страница 72: ...72 P a g e cascadesciences com Tel 503 847 9047 Continued next page...
Страница 95: ...95 P a g e cascadesciences com Tel 503 847 9047 End of Procedure...
Страница 102: ...102 P a g e cascadesciences com Tel 503 847 9047 Note See the next page for how to begin heating the oven chamber...
Страница 104: ...104 P a g e cascadesciences com Tel 503 847 9047 End of Procedure...
Страница 111: ...111 P a g e cascadesciences com Tel 503 847 9047 Tap the red timer Stop button to clear the Timer Finished message...
Страница 116: ...116 P a g e cascadesciences com Tel 503 847 9047 End of Procedure...
Страница 123: ...123 P a g e cascadesciences com Tel 503 847 9047 Continued next page...
Страница 125: ...125 P a g e cascadesciences com Tel 503 847 9047 Continued next page...
Страница 127: ...127 P a g e cascadesciences com Tel 503 847 9047 End of Procedure...
Страница 128: ...128 P a g e cascadesciences com Tel 503 847 9047 MAINTENANCE...
Страница 130: ...130 P a g e cascadesciences com Tel 503 847 9047 REPLACEMENT PARTS...
Страница 132: ......