MAINTENANCE
6. Heat up and stabilization period
: The oven shelving temperature must be stable at
temperature in a vacuum environment to perform an accurate calibration.
•
The temperature is considered stabilized when the oven chamber has operated at your
calibration temperature for at least 1 hour with no fluctuations greater than the specified
temperature stability of the oven.
Oven Chamber Heat Up and Stabilization Phases
Suggested Calibration Procedure
1.
Compare the reference device and oven temperature display readings.
•
If the readings are the same, or the difference between the two falls within the acceptable
range of your protocol, the display is accurately showing the chamber shelf temperature.
The Temperature Calibration procedure is now complete.
•
14T14T14T
If the difference falls outside of your protocol range, advance to the next step
14T14T14T
.
40°C
Start
Required Stability
Period 1 Hour
Mi i
Begin Calibration
Heat Up to 40°C:
~60 Minutes
Fluctuations
(Exaggerated
)
Reference
Reference Device
Unit in Calibration
Calibration Required
Содержание CVO-5-EX Series
Страница 1: ...Flammable Solvent Ovens Installation Operation Manual CVO 5 EX Oven Series...
Страница 6: ...6 P a g e cascadesciences com Tel 503 847 9047 Special Conditions Of Use 131...
Страница 7: ...7 P a g e cascadesciences com Tel 503 847 9047 TABLE OF CONTENTS...
Страница 11: ...11 P a g e cascadesciences com Tel 503 847 9047 SAFETY CERTS AND STATEMENTS...
Страница 17: ...17 P a g e cascadesciences com Tel 503 847 9047 SPECIFICATIONS...
Страница 22: ...22 P a g e cascadesciences com Tel 503 847 9047...
Страница 24: ...24 P a g e cascadesciences com Tel 503 847 9047 These groups include propane and n butane...
Страница 29: ...29 P a g e cascadesciences com Tel 503 847 9047 Continued next page...
Страница 37: ...37 P a g e cascadesciences com Tel 503 847 9047 INTRODUCTION...
Страница 39: ...39 P a g e cascadesciences com Tel 503 847 9047...
Страница 44: ...44 P a g e cascadesciences com Tel 503 847 9047 RECEIVING...
Страница 51: ...51 P a g e cascadesciences com Tel 503 847 9047 Continued next page...
Страница 54: ...54 P a g e cascadesciences com Tel 503 847 9047 Continued next page...
Страница 61: ...61 P a g e cascadesciences com Tel 503 847 9047 End of Procedure...
Страница 65: ...65 P a g e cascadesciences com Tel 503 847 9047 INSTALLATION...
Страница 67: ...67 P a g e cascadesciences com Tel 503 847 9047 SYMBOLS...
Страница 69: ...69 P a g e cascadesciences com Tel 503 847 9047 Continued next page...
Страница 72: ...72 P a g e cascadesciences com Tel 503 847 9047 Continued next page...
Страница 95: ...95 P a g e cascadesciences com Tel 503 847 9047 End of Procedure...
Страница 102: ...102 P a g e cascadesciences com Tel 503 847 9047 Note See the next page for how to begin heating the oven chamber...
Страница 104: ...104 P a g e cascadesciences com Tel 503 847 9047 End of Procedure...
Страница 111: ...111 P a g e cascadesciences com Tel 503 847 9047 Tap the red timer Stop button to clear the Timer Finished message...
Страница 116: ...116 P a g e cascadesciences com Tel 503 847 9047 End of Procedure...
Страница 123: ...123 P a g e cascadesciences com Tel 503 847 9047 Continued next page...
Страница 125: ...125 P a g e cascadesciences com Tel 503 847 9047 Continued next page...
Страница 127: ...127 P a g e cascadesciences com Tel 503 847 9047 End of Procedure...
Страница 128: ...128 P a g e cascadesciences com Tel 503 847 9047 MAINTENANCE...
Страница 130: ...130 P a g e cascadesciences com Tel 503 847 9047 REPLACEMENT PARTS...
Страница 132: ......