USER MAINTENANCE
L
eak Verification, Y/N?
Verify that the oven is leaking and not experiencing outgassing from residual contamination.
1.
If oven is hot, allow oven to cool to ambient temperature.
2.
Evacuate the chamber to the lowest achievable pressure.
3.
Write down the pressure displayed on the Vacuum Gauge as a positive number.
4.
Isolate the chamber by closing the vacuum valve.
5.
Monitor the vacuum gauge for 1 hour. If the pressure on the gauge increases, this may indicate a
significant leak.
6.
Monitor the gauge for another hour.
•
If the pressure level stabilizes, the initial rise in pressure may be caused by residual
material in the chamber outgassing, increasing the pressure.
•
If the pressure level continues to increase, record the pressure displayed on the Vacuum
Gauge at the end of the second hour. Include these findings in the Diagnostic Data Log.
Содержание CVO-2-HT
Страница 1: ...Vacuum Ovens CVO 2 HT CVO 5 HT Installation Operation Manual...
Страница 2: ...2 P a g e cascadesciences com Tel 503 847 9047 Pictured on front cover left to right CVO 2 HT CVO 5 HT...
Страница 6: ...6 P a g e cascadesciences com Tel 503 847 9047 TABLE OF CONTENTS...
Страница 8: ...8 P a g e cascadesciences com Tel 503 847 9047 CERTIFICATIONS...
Страница 12: ...12 P a g e cascadesciences com Tel 503 847 9047 UNIT SPECIFICATIONS...
Страница 16: ...16 P a g e cascadesciences com Tel 503 847 9047 INTRODUCTION...
Страница 22: ...22 P a g e cascadesciences com Tel 503 847 9047 RECEIVING YOUR OVEN...
Страница 32: ...32 P a g e cascadesciences com Tel 503 847 9047 GRAPHIC SYMBOLS...
Страница 71: ...71 P a g e cascadesciences com Tel 503 847 9047 REPLACEMENT PARTS...
Страница 72: ......