59
|
P a g e
cascadesciences.com
|
Tel. 503 847-9047
USER MAINTENANCE
Note:
Does the car actually have gas in the tank? Have you physically verified the computer is
plugged in? Yes, we are going ask some very basic questions. Please bear with us.
Methodical verifications and the elimination of potential causes of failure are often the
quickest means of getting a unit back into operation.
Unit Performance Specifications
Verify the following items to ensure a fault in the unit is preventing it from achieving its specified
performance levels.
Standard Ventilation Spacing Requirements
Verify there is sufficient spacing around the oven for ventilation or cooling. Insufficient spacing can
adversely impact temperature performance.
•
A minimum of
6 inches (152 mm)
clearance is required between the sides of the oven and any
walls or partitions.
•
A minimum of
12 inches (305 mm)
between the top of the oven and overhead cover.
•
At least
12 inches (305 mm)
between the back of the oven and any partitions or objects.
Operating Temperature Range of the Unit
Verify that your constant temperature set point or the set points of your heating profile all fall within
the operating range of the oven.
•
Find the Temperature Range specifications for the oven in the Unit Specifications chapter of the
Operation Manual, page 9.
Содержание CVO-2-HT
Страница 1: ...Vacuum Ovens CVO 2 HT CVO 5 HT Installation Operation Manual...
Страница 2: ...2 P a g e cascadesciences com Tel 503 847 9047 Pictured on front cover left to right CVO 2 HT CVO 5 HT...
Страница 6: ...6 P a g e cascadesciences com Tel 503 847 9047 TABLE OF CONTENTS...
Страница 8: ...8 P a g e cascadesciences com Tel 503 847 9047 CERTIFICATIONS...
Страница 12: ...12 P a g e cascadesciences com Tel 503 847 9047 UNIT SPECIFICATIONS...
Страница 16: ...16 P a g e cascadesciences com Tel 503 847 9047 INTRODUCTION...
Страница 22: ...22 P a g e cascadesciences com Tel 503 847 9047 RECEIVING YOUR OVEN...
Страница 32: ...32 P a g e cascadesciences com Tel 503 847 9047 GRAPHIC SYMBOLS...
Страница 71: ...71 P a g e cascadesciences com Tel 503 847 9047 REPLACEMENT PARTS...
Страница 72: ......