65
|
P a g e
cascadesciences.com
|
Tel. 503 847-9047
USER MAINTENANCE
Before Starting
1.
The unit must be connected to a power source that meets the requirements in the Installation chapter
(page 25) and turned on.
2.
Do not heat the oven.
The oven must remain at ambient temperature for this procedure.
3.
The oven chamber must be empty, sealed, and under full achievable vacuum levels. See the
Evacuating the Oven Chamber
entry on page 42.
4.
Check the primary chamber gasket for damage
. This is the gasket mounted either on the chamber
liner or the door that seals the oven chamber when the door is closed.
Look for:
•
Cuts or nicks on the gasket caused by removing shelves or samples from the
chamber.
•
Cracking, brittleness, or loss of elasticity.
•
Discoloration of the gasket.
•
Nicks or other damage on the surface the gasket seals against.
Vacuum Gauge
Содержание CVO-2-HT
Страница 1: ...Vacuum Ovens CVO 2 HT CVO 5 HT Installation Operation Manual...
Страница 2: ...2 P a g e cascadesciences com Tel 503 847 9047 Pictured on front cover left to right CVO 2 HT CVO 5 HT...
Страница 6: ...6 P a g e cascadesciences com Tel 503 847 9047 TABLE OF CONTENTS...
Страница 8: ...8 P a g e cascadesciences com Tel 503 847 9047 CERTIFICATIONS...
Страница 12: ...12 P a g e cascadesciences com Tel 503 847 9047 UNIT SPECIFICATIONS...
Страница 16: ...16 P a g e cascadesciences com Tel 503 847 9047 INTRODUCTION...
Страница 22: ...22 P a g e cascadesciences com Tel 503 847 9047 RECEIVING YOUR OVEN...
Страница 32: ...32 P a g e cascadesciences com Tel 503 847 9047 GRAPHIC SYMBOLS...
Страница 71: ...71 P a g e cascadesciences com Tel 503 847 9047 REPLACEMENT PARTS...
Страница 72: ......