53
|
P a g e
cascadesciences.com
|
Tel. 503 847-9047
USER MAINTENANCE
Continued on next page
Calibration continued
3
Unlock the controller.
a.
See the Unlocking procedure on page 55.
Note:
The temperature controller must be unlocked in order to
access the Operations menu and enter a calibration offset.
4
Navigate to the Operations menu
after
unlocking the
controller.
a.
Press and hold both the
Up
and
Down
Arrow buttons
simultaneously for approximately 5 seconds.
b.
Release the buttons when “A1” appears on the top
display line and “oPEr” appears in the mid display line.
Operations Menu
5
Advance through the Operations menu options to the
Temperature Calibration offset parameter.
Push the green
Advance button
repeatedly until “i.CA”
appears in the green mid display line and a number value in
the red top line.
6
Adjust the number value in the top display line to match the
offset value from step 2, using the arrow buttons.
7
Save the calibration offset and return to the homepage.
a.
Push the Reset button 3 times so the display shows the
homepage.
•
The oven will begin to heat or passively cool to reach the
current set point using the offset display value for the current
temperature.
x2
x3
Содержание CVO-2-HT
Страница 1: ...Vacuum Ovens CVO 2 HT CVO 5 HT Installation Operation Manual...
Страница 2: ...2 P a g e cascadesciences com Tel 503 847 9047 Pictured on front cover left to right CVO 2 HT CVO 5 HT...
Страница 6: ...6 P a g e cascadesciences com Tel 503 847 9047 TABLE OF CONTENTS...
Страница 8: ...8 P a g e cascadesciences com Tel 503 847 9047 CERTIFICATIONS...
Страница 12: ...12 P a g e cascadesciences com Tel 503 847 9047 UNIT SPECIFICATIONS...
Страница 16: ...16 P a g e cascadesciences com Tel 503 847 9047 INTRODUCTION...
Страница 22: ...22 P a g e cascadesciences com Tel 503 847 9047 RECEIVING YOUR OVEN...
Страница 32: ...32 P a g e cascadesciences com Tel 503 847 9047 GRAPHIC SYMBOLS...
Страница 71: ...71 P a g e cascadesciences com Tel 503 847 9047 REPLACEMENT PARTS...
Страница 72: ......