INSTALLATION
L
EVELING
Install the 4 leveling feet in the 4 corner holes in the bottom of the oven.
The oven must be level and stable for safe operation.
Note:
To prevent damage when moving the unit, turn all four leveling feet so that the leg of each
foot sits inside the unit.
I
NSTALL THE
O
VEN
Place the unit in a workspace location that meets the criteria discussed in the previous entries of the
Installation section.
•
Verify that the oven stands level and does not rock. Adjust the leveling feet as needed.
I
NSTALLATION
C
LEANING
The unit was cleaned at the factory, but not sterilized. It may have been exposed to contaminants en route
during shipping.
•
Remove all wrappings and coverings from shelving prior to cleaning and installation.
•
Do not clean with deionized water.
•
topic in the User Maintenance section (see page 45) for
more information on how to clean the oven chamber prior to putting the unit into operation.
Содержание CVO-10
Страница 1: ...Vacuum Purge Ovens Installation Operation Manual CVO 5 CVO 5 2 CVO 10 ...
Страница 4: ...4 P a g e cascadesciences com Tel 503 847 9047 REPLACEMENT PARTS 53 ...
Страница 6: ...6 P a g e cascadesciences com Tel 503 847 9047 CERTIFICATIONS ...
Страница 10: ...10 P a g e cascadesciences com Tel 503 847 9047 UNIT SPECIFICATIONS ...
Страница 18: ...18 P a g e cascadesciences com Tel 503 847 9047 RECEIVING YOUR OVEN ...
Страница 28: ...28 P a g e cascadesciences com Tel 503 847 9047 INSTALLATION ...
Страница 30: ...30 P a g e cascadesciences com Tel 503 847 9047 GRAPHIC SYMBOLS ...
Страница 52: ...52 P a g e cascadesciences com Tel 503 847 9047 USER MAINTENANCE ...
Страница 55: ...55 P a g e cascadesciences com Tel 503 847 9047 REPLACEMENT PARTS ...
Страница 56: ......