50 |
P a g e
cascadesciences.com
|
Tel. 503 847-9047
USER MAINTENANCE
Procedure continued on next page
Calibration continued
3
jump to the temperature setpoint menu
a.
Press the blue setpoint button on the
display screen homepage.
•
The display will jump to the temperature
setpoint page.
The temperature setpoint
menu
4
Place the setpoint page into its calibration mode.
a.
Press and hold the hidden button (blank space)
on the top right of the screen.
•
The display changes to “Calibrate To”
5
Use the keypad to enter the temperature shown on your
reference device. This will match the oven display to the
reference device reading.
Reference Device
6
Save the calibration match, then return to the homepage.
a.
Press the Save Button
b.
Press the Home Button
•
The oven will now heat or passively cool to
achieve the setpoint temperature with the
display correction.
Reference Device
Setpoint
Hidden
Button
Содержание CVO-10
Страница 1: ...Vacuum Purge Ovens Installation Operation Manual CVO 5 CVO 5 2 CVO 10 ...
Страница 4: ...4 P a g e cascadesciences com Tel 503 847 9047 REPLACEMENT PARTS 53 ...
Страница 6: ...6 P a g e cascadesciences com Tel 503 847 9047 CERTIFICATIONS ...
Страница 10: ...10 P a g e cascadesciences com Tel 503 847 9047 UNIT SPECIFICATIONS ...
Страница 18: ...18 P a g e cascadesciences com Tel 503 847 9047 RECEIVING YOUR OVEN ...
Страница 28: ...28 P a g e cascadesciences com Tel 503 847 9047 INSTALLATION ...
Страница 30: ...30 P a g e cascadesciences com Tel 503 847 9047 GRAPHIC SYMBOLS ...
Страница 52: ...52 P a g e cascadesciences com Tel 503 847 9047 USER MAINTENANCE ...
Страница 55: ...55 P a g e cascadesciences com Tel 503 847 9047 REPLACEMENT PARTS ...
Страница 56: ......