49 |
P a g e
cascadesciences.com
|
Tel. 503 847-9047
Begin Calibration
USER MAINTENANCE
5. Heat up and stabilization period
: The oven temperature must be stable at temperature and under
vacuum in order to perform an accurate calibration.
•
The temperature is considered stabilized when the oven chamber has operated at your
calibration temperature for at least 30 minutes with no fluctuations greater than
±0.4°F
.
Oven Chamber Heat Up and Stabilization Phases
Continued on next page
Suggested Calibration Procedure
1
Once the chamber has stabilized with no fluctuations, compare the
reference temperature device and chamber temperature display
readings.
•
If the readings are the same, or the difference
between the two falls within the acceptable range of
your protocol, the display is accurately showing the
chamber temperature.
The Temperature Calibration
procedure is now complete
.
-Or-
•
See step 2 if a difference falls outside the acceptable
range of your protocol.
Reference Device
2
The display requires calibration. Advance to step 3.
Reference Device
Set Temperature
105°F
Start
30 Minute Minimum
Stability Period
Heat Up to 105°F:
~120 Minutes
Fluctuations
(Exaggerated)
Содержание CVO-10
Страница 1: ...Vacuum Purge Ovens Installation Operation Manual CVO 5 CVO 5 2 CVO 10 ...
Страница 4: ...4 P a g e cascadesciences com Tel 503 847 9047 REPLACEMENT PARTS 53 ...
Страница 6: ...6 P a g e cascadesciences com Tel 503 847 9047 CERTIFICATIONS ...
Страница 10: ...10 P a g e cascadesciences com Tel 503 847 9047 UNIT SPECIFICATIONS ...
Страница 18: ...18 P a g e cascadesciences com Tel 503 847 9047 RECEIVING YOUR OVEN ...
Страница 28: ...28 P a g e cascadesciences com Tel 503 847 9047 INSTALLATION ...
Страница 30: ...30 P a g e cascadesciences com Tel 503 847 9047 GRAPHIC SYMBOLS ...
Страница 52: ...52 P a g e cascadesciences com Tel 503 847 9047 USER MAINTENANCE ...
Страница 55: ...55 P a g e cascadesciences com Tel 503 847 9047 REPLACEMENT PARTS ...
Страница 56: ......