Reference Manual V1.10
Software Versions 3.xx
X300-621-110
Page 53
17.
Error Messages
A number of error messages may be displayed to warn of operation outside of the acceptable limits.
These messages are described below. Short messages (XXXXX) will appear as a single message
on the display. Longer messages (XXXXX) (YYYYY) will appear on the display in two parts, first the
(XXXXX) part, then the (YYYYY) part.
17.1. Weighing Errors
These messages show status messages or errors that may occur during normal weighing operation.
Error
Description
Resolution
(U - - - - -)
The weight is below the minimum allowable
weight reading.
Increase the weight or
decrease the minimum
allowable weight reading.
(O - - - - -)
The weight is above the maximum allowable
weight reading.
Warning - overloading may damage
mechanical scale elements.
Check the condition of
load cell connections.
Check for damaged load
cell.
(ZERO)
(ERROR)
The weight reading is beyond the limit set for
Zero operation. The operation of the
<ZERO>
key is limited in the setup during
installation. The indicator cannot be Zeroed
at this weight.
Increase the Zero Range
(Z.RANGE) or use the
<TARE>
key instead.
(STABLE)
(ERROR)
Scale motion has prevented a
<ZERO>
or
<TARE>
operation from occurring on
command.
Try the operation again
once the scale is stable.