ALARMS
Alarm System Overview
The chamber control system is equipped with an alarm system that constantly monitors
H2O2 Levels and Power Loss to ensure the user is notified if the chamber goes into an
alarm condition. Notification occurs via an alarm pop-up window and a buzzer. If the
optional remote alarm contacts are present, in an alarm condition, the dry contacts will
change state.
The following alarm messages may be displayed:
H2O2 Level High; high level of H2O2 inside of chamber
Power Loss;Cycle Aborted
In the event an alarm occurs, the alarm indicator will appear on the status bar and an
audible alarm pop-up window will automatically appear.
In the event of a power loss,or turned off at power switch the unit will
automatically restart when the power is restored or the switch is on.
The Sterilization Cycle will run as if the cycle was aborted. Therefore
you cannot open the door until the Inactivation Cycle is complete. This
is a built in safety feature.
When the cycle is complete the Done button will appear on the screen
You will hear an audible “click” from the Door Interlock then the door
can be opened.