7800-25 Series Operations Manual
Rev B 11-10-2020
Page 35 of 56
Calibrating Temperature
If temperature calibration is needed, the following steps can be taken:
Locate the reference instrument’s temperature sensor in close proximity to the
cabinet’s geometric center. Be sure that the stabilization times described earlier have
been satisfied prior to performing calibration.
At the calibrate screen, press the
(Temperature Calibrate) button.
Enter the temperature offset by using the keypad and pressing
(Enter) when
complete.
A positive value will move the temperature ‘up’ and a negative value ‘down’. Press the
‘home’ button and verify the proper temperature is displayed.
Temperature Offset
Calibrate Button
Содержание 7800-25
Страница 1: ...Sterilization Chamber Model 7800 25 ...