7800-25 Series Operations Manual
Rev B 11-10-2020
Page 33 of 56
CALIBRATION
The temperature, humidity and H2O2 heater systems require calibration. CARON
recommends an annual calibration check of each system. Before making a calibration
adjustment, allow the cabinet to stabilize a minimum of 12 hours from a power off
condition with the door open and must be done with no H2O2 vapors present.
If you do not have the appropriate reference instruments to perform calibration, contact
CARON’s service department for on-site calibration at www.caronproducts.com Caron
also provides validation services which ensure that the unit is functioning properly
according to IQ, OQ and PQ protocols which satisfy FDA guidelines for qualification
verification of equipment.
The Calibration Screen
To get to the calibration screen from the home page:
Press the
(Menu) button.
Be sure that all reference instruments are
calibrated to an appropriate standard.
Menu
Button