7800-25 Series Operations Manual
Rev B 11-10-2020
Page 14 of 56
Connect each chamber to a grounded circuit.
Failure to do so could result in electrical shock.
Connecting Electrical Power
The unit requires a dedicated electrical outlet. See table below for model specific power
required and connection.
Model #
Power Requirements
Plug Connection
-1
115V, 60Hz, 12A FLA
NEMA 5-15
-2
230V, 60Hz, 8A FLA
NEMA 6-15
-3
230V, 50Hz, 8A FLA
CEE 7/7
When the required electrical connection is available, plug the provided power cord into
the unit and the electrical outlet.
The mains power supply cord must meet the requirements listed
above. The use of an inadequate mains power supply cord
could result in equipment failure or personal harm to the user.
In the event of a power outage the unit will automatically restart
when the power is restored. There are built safeties that will not
allow the door to be opened until chamber conditions are at a
safe level.