CALIBRATION
The temperature and optional humidity system can all be calibrated as necessary.
CARON recommends an annual calibration check of each system. Before making a
calibration adjustment, allow the cabinet to stabilize a minimum of 12 hours from a power
off condition. If the unit has been in operation, allow a minimum of 3 hours of stable
operation at all setpoints.
If you do not have the appropriate reference instruments to perform calibration, contact
CARON’s service department for on-site calibration at
www.caronproducts.com
Caron
also provides validation services which ensure that the unit is functioning properly
according to IQ, OQ and PQ protocols which satisfy FDA guidelines for qualification
verification of equipment.
The Calibration Screen
To get to the calibration screen from the home page:
Press the
(Settings) button.
Be sure that all reference instruments are
calibrated to an appropriate standard.
Settings
Button
Содержание 7340-25
Страница 1: ......
Страница 80: ...7340_7350 25 33 Series Operations Manual Rev J 3 31 21 Page 80 of 92 ELECTRICAL SCHEMATICS ...
Страница 81: ...7340_7350 25 33 Series Operations Manual Rev J 3 31 21 Page 81 of 92 ...
Страница 82: ...7340_7350 25 33 Series Operations Manual Rev J 3 31 21 Page 82 of 92 ...
Страница 83: ...7340_7350 25 33 Series Operations Manual Rev J 3 31 21 Page 83 of 92 ...
Страница 84: ...7340_7350 25 33 Series Operations Manual Rev J 3 31 21 Page 84 of 92 ...
Страница 85: ...7340_7350 25 33 Series Operations Manual Rev J 3 31 21 Page 85 of 92 ...
Страница 86: ...7340_7350 25 33 Series Operations Manual Rev J 3 31 21 Page 86 of 92 ...
Страница 87: ...7340_7350 25 33 Series Operations Manual Rev J 3 31 21 Page 87 of 92 ...
Страница 88: ...7340_7350 25 33 Series Operations Manual Rev J 3 31 21 Page 88 of 92 ...