7340_7350-25-33 Series Operations Manual
Rev J 3-31-21
Page 77 of 92
PREVENTATIVE MAINTENANCE
The CARON chamber has been robustly designed to minimize performance problems.
However, regular maintenance is very important for continuous trouble free operation.
As a general rule, CARON recommends an annual calibration check of the temperature
and humidity systems. CARON offers a full range of on-site calibration and validation
services. We also offer preventative maintenance contracts on our equipment. Contact
our Service department for details at 740-373-6809 or visit us on the web at
www.caronproducts.com
.
Recommended Daily Maintenance Checks
Check the Temperature and Humidity displays versus setpoints.
Check for and correct any alarm condition.
Recommended Monthly Maintenance Checks
Check to ensure the drain in the bottom of the unit is draining properly.
Check front air intake filter. If the filter is dirty replace it with Caron Preventative
Maintenance PM Kit. Washing the filter will result in poor performance.
Recommended Annual Maintenance Checks
Replace humidity atomizer nozzle (HUMD311-1,-4 option) (see Replacement Parts
section).
Disinfect all interior surfaces with a general purpose laboratory cleaning agent.
Perform a complete calibration of the temperature and humidity systems.
A full validation is recommended for GMP facilities each time a unit is installed, moved
or undergoes significant repair. Contact CARON’s service department to schedule
on-site validation.
Содержание 7340-25
Страница 1: ......
Страница 80: ...7340_7350 25 33 Series Operations Manual Rev J 3 31 21 Page 80 of 92 ELECTRICAL SCHEMATICS ...
Страница 81: ...7340_7350 25 33 Series Operations Manual Rev J 3 31 21 Page 81 of 92 ...
Страница 82: ...7340_7350 25 33 Series Operations Manual Rev J 3 31 21 Page 82 of 92 ...
Страница 83: ...7340_7350 25 33 Series Operations Manual Rev J 3 31 21 Page 83 of 92 ...
Страница 84: ...7340_7350 25 33 Series Operations Manual Rev J 3 31 21 Page 84 of 92 ...
Страница 85: ...7340_7350 25 33 Series Operations Manual Rev J 3 31 21 Page 85 of 92 ...
Страница 86: ...7340_7350 25 33 Series Operations Manual Rev J 3 31 21 Page 86 of 92 ...
Страница 87: ...7340_7350 25 33 Series Operations Manual Rev J 3 31 21 Page 87 of 92 ...
Страница 88: ...7340_7350 25 33 Series Operations Manual Rev J 3 31 21 Page 88 of 92 ...