7.3.2 Checking the maintenance data in the controller’s maintenance view
With the device running, press and hold
down the lock and information signal re-
set button for three (3) seconds to access
the maintenance view.
In the view, you can check the device
software version, (
Firmware
), serial
number (
S/N
), pump operation hours (
Us-
age h (p)
), date of device commissioning
(
Device
), date of battery commissioning
(
Battery
) and the date of the following
scheduled maintenance (
Service
).
To change from one tab to another in the
maintenance view, press the information
signal reset button.
To return from the maintenance view to
the normal operating mode, press and
hold down the information signal reset
button for three (3) seconds.
After returning from the maintenance
view, the device checks the pressure val-
ues of the mattress system.
7.4 Storage and transport
Decommissioning the mattress system
1. Press the standby button to stop the device
2. Disconnect the power cable by pressing
operation and then disconnect the device’s
the yellow button and pulling the cable
power plug from the mains.
out.
1/2
Firmware:
S/N:
Usage h (p):
2/2
Device:
Battery:
Service:
3 s.
3 s.
3 s.