(CS 2400P) reduced 60% compared to the adults based on the literature.
National DRLs have been typically set at a rounded 75th percentile of the dose distribution.
The values presented in Table are proposed as new NDRLs for the UK.
X-ray type
Patient size
Proposed NDRL
Intra oral
Adult mandibular moral
1.2mGy
Child mandibular moral
0.7mGy
And We got the data using Portable Dental X-ray (CS 2400P). These exposed dose
values are more safer than requirements of NDRLs for the UK in adults and children.
Dose Table (70 kVp, 3.5 mA, FOV: Ø 6 cm, SSD 200 mm, at Al 6 mm)
t (s)
Dose (μGy)
0.06
41
0.10
68
0.14
95
0.20
134
0.34
221
0.5
320
0.6
377
0.8
487
1.0
595
Table below is information about used equipment for test.
Test Equipment
Instrument
Manufact
urer
Model
S/N
Dose Meter
Ray-safe
X2 Base Unit
225575
•
Strategy of reduction of X-ray for pediatric patients
According to the prior results the change over age and gender for Mandibular Angle
Breadth is 3% at maximum. And the BMI which related to body shape have change over
age and gender of 6% at maximum.
Therefore, we simplified the maximum change over gender, age, body shape as 6% and
we determined the range for exposure time.
In case of child, according to NDRL report*, the child to adult DAP ratio at 50th percentile
at mandibular molar intra-oral radiograph, panoramic radiograph, CBCT is 0.6, 0.7 and 0.6
Содержание Nanoray CS 2400P
Страница 1: ...Battery Free Portable Dental X ray USER MANUAL Doc UM 02FC 01 Rev 5 2021 01 29...
Страница 8: ...page 8 62 1 4 2 Labels and label positions Device labels 55x80mm NR F350 and CS 2400P...
Страница 10: ...page 10 62 Adaptor Label 38 X 90mm Warning Label 50 X 20mm...
Страница 11: ...page 11 62 Caution Label 60 X 20mm...