page
15
/
62
Especially for pregnant women, it is recommended to use personal dosimetry and a
lead apron
- The vertical significant zone of occupancy measures 60 cm X 200 cm, while the
horizontal significant zone of occupancy is 60 cm X 60 cm.
- The Backscatter Shield : In addition to the radiation shielded cone, the back scatter
shield provides additional protection to the operator.
- DO NOT attempt to remove the backscatter shield. Attempting to do so will result in
damage to your device.
- The equipment can be used at an altitude of less than 3,000 meters.
- WARNING: Use of this equipment adjacent to or stacked with other equipment should
be avoided because it could result in improper operation. If such use is necessary, this
equipment and the other equipment should be observed to verify that they are
operating normally.
- WARNING: Use of accessories, transducers and cables other than those specified or
provided by the manufacturer of this equipment could result in increased
electromagnetic emissions or decreased electromagnetic immunity of this equipment
and result in improper operation.
- WARNING: Portable RF communications equipment (including peripherals such as
antenna cables and external antennas) should be used no closer than 30 cm (12
inches) to any part of the [ME EQUIPMENT or ME SYSTEM], including cables specified
by the manufacturer. Otherwise, degradation of the performance of this equipment
could result.
- WARNING: Failure to use this equipment in the specified type of shielded location
could result in degradation of the performance of this equipment, interference with other
equipment or interference with radio services.
- Metal substances or liquids shall not come into contact with the charger pogo-pins of
the product and the charging cradle.
Radiation Safety
-
This equipment should be operated by a trained and qualified Dentist or a Dental
hygienist in a controlled environment.
-
All users and patients should wear protective equipment, such as a lead apron
which is covered until collar.
When using the equipment, it is recommended that all users comply
with the following radiation safety guidance for the safety of the users
and the patients.
Содержание Nanoray CS 2400P
Страница 1: ...Battery Free Portable Dental X ray USER MANUAL Doc UM 02FC 01 Rev 5 2021 01 29...
Страница 8: ...page 8 62 1 4 2 Labels and label positions Device labels 55x80mm NR F350 and CS 2400P...
Страница 10: ...page 10 62 Adaptor Label 38 X 90mm Warning Label 50 X 20mm...
Страница 11: ...page 11 62 Caution Label 60 X 20mm...