62
2) X-ray Generator
Item
Specifications
High Voltage
Generator
(Assembly)
Model
NRF-350-G1
Rated output power
Max. 0.21 kW
Duty Cycle
1:90
(Exposure time: Interval time)
Cooling Protection
Thermistor ≥
70 °C
Type
Inverter Type
Tube Voltage
65-75kV
Tube Current
2.8 - 4.2 mA
X-ray Tube
Manufacturer
Hangzhou Kailong Medical
Instrument Co., Ltd
Model
KL-11-0.4-70
Focal spot size
0.4 mm (IEC 60336)
Anode heat contents
4.5 kJ
Maximum Anode Heat
Dissipation
110 W
Target Material
Tungsten
Target Angle
12
°
Inherent Filtration
Min. 0.8mm Al Eq @ 75kV
X-ray Coverage
70 mm at SID 200 mm
Tube Voltage
70 kV
Tube Current
Max. 12.0 mA