2.1
General Safety Guidelines
1) General Safety Guidelines
- This product requires a longer break time than the exposure time after an X-ray
exposure. The minimum duty cycle rating (the relationship between duration and
frequency of exposures) is 1:90.
Duration
0.06s
0.32s
0.56s
1s
Cycle
Every 5.4s
Every 28.8s
Every 50.4s
Every 90s
- This product is designed and manufactured in order to ensure the maximum safety.
Operation and Maintenance must be in accordance with the instructions contained in
this manual.
- This product must only be operated by legally qualified.(Dentist or Dental hygienist)
- Observe all the local fire regulations keep a fire extinguisher near the product at all
times.
- Maintenance and services for the product should be taken by qualified service
personnel according to procedures and preventive maintenance schedules.
- Always turn off the product before cleaning.
- The device uses beam limiting device(rectangular cover) for protection against stray
radiation
2) Limit product temperature
- In order to protect the over temperature of the product, exposure shall not be
conducted if the generator temperature is higher than the set value.
- Display shows the measured temperature. (Ex display: 45.0
℃
)
- Before the generator temperature goes down, the downtime of the product is required
for at least 1 hour.
3) For self-protection functions, cooling time of about 5 minutes is required if the downtime
is insufficient.
- The downtime required for the display is displayed, and it cannot be exposure. (Ex
display : "300s")
- Self-protection function: Function that prevents temporary use of the product to
compensate for insufficient cooling time.
2. Safety Instructions
Содержание Nanoray CS 2400P
Страница 1: ...Battery Free Portable Dental X ray USER MANUAL Doc UM 02FC 01 Rev 5 2021 01 29...
Страница 8: ...page 8 62 1 4 2 Labels and label positions Device labels 55x80mm NR F350 and CS 2400P...
Страница 10: ...page 10 62 Adaptor Label 38 X 90mm Warning Label 50 X 20mm...
Страница 11: ...page 11 62 Caution Label 60 X 20mm...