page
52
/
62
Guidance and manufacturer’s declaration - electromagnetic immunity
The CS 2400P model device is intended for use in the electromagnetic environment specified
below.
The customer or the user of the CS 2400P model device should assure that it is used in such an
environment
IMMUNITY test
IEC 60601-1-2 test level
Electromagnetic environments
Conducted
disturbances
induced by RF
fields
IEC 61000-4-6
3 V 0,15 MHz to 80 MHz,
6 V in ISM bands between 0,15 MHz
and 80 MHz, 80 % AM at 1 kHz
(Professional healthcare facility
environment)
3 V, 0,15 MHz to 80 MHz
6 V in ISM and amateur radio bands
between 0,15 MHz and 80 MHz, 80 %
AM at 1 kHz
(Home healthcare environment)
Floors should be wood, concrete
or ceramic tile If floors are
covered with synthetic material,
the relative humidity should be at
least 30 %
Radiated RF EM
fields
IEC 61000-4-3
3 V/m, 80 MHz to 2,7 GHz, 80 % AM at
1 kHz
(Professional healthcare facility
environment)
10 V/m, 80 MHz to 2,7 GHz, 80 % AM
at 1 kHz
(Home healthcare environment)
Portable and mobile RF
communications equipment
should be used closer to any part
of the CS 2400P model device,
including cables, than the
recommended separation
distance calculated from the
equation applicable to the
frequency of the transmitter.
Recommended separation
distance
d
= 1.2
√
P
d
= 1.2
√
P
, 80 MHz to 800 MHz
d
= 2.3
√
P
, 800 MHz to 2.7 GHz
Where
P
is the maximum output
power rating of the transmitter in
watts (W) according to the
transmitter manufacturer and
d
is
the recommended separation
distance in meters (m)
Field strengths from fixed RF
transmitters, as determined by an
electromagnetic site, survey,
a
should be less than the
compliance level in each
frequency range.
b
Interference
may occur in the vicinity of
equipment marked with the
following symbol :
Содержание Nanoray CS 2400P
Страница 1: ...Battery Free Portable Dental X ray USER MANUAL Doc UM 02FC 01 Rev 5 2021 01 29...
Страница 8: ...page 8 62 1 4 2 Labels and label positions Device labels 55x80mm NR F350 and CS 2400P...
Страница 10: ...page 10 62 Adaptor Label 38 X 90mm Warning Label 50 X 20mm...