page
29
/
62
4.11
Interfaces between the software system and other systems
User Interface
Function
Interface specification
Button
-
Program start, end
-
Teeth selection
-
Select the exposure object
-
Select the type of exposure
-
Initialize exposure time
-
Administrator mode entry and exit
-
Change administrator mode
-
Select the Lock/unlock mode
6ea button (Beside LCD
display)
Wheel
-
Select the exposure time
-
Change administrator mode setting
value
1ea wheel
Exposure
Switch
-
Exposure
1ea Exposure Switch
LCD display
-
Display the start status of the system
-
Display the current status of the system
(charge, exposure, battery level,
exposure object selection mode,
exposure time and Charge state, tooth
selection)
3.5 inch
(Out size dimension : 64 X
46mm)
(Viewing Area Dimension :
60X 41mm)
Cradle
-
Charge
-
Charging status LED indication
1 Cradle
Buzzer
-
Alarm sound output
1 Buzzer
External
exposure
switch
-
Exposure
1ea External exposure
Switch
4.12
Main functional elements and functionality
No.
Contents
Function
1
System On/Off function
1)
Display the start status of the system
2)
Display the shutdown status of the system
2
Information of System
status and
Progress message
Display of the current status of the system (charging,
exposure, battery level, exposure object selection,
exposure time and charge status, tooth selection, etc.)
Содержание Nanoray CS 2400P
Страница 1: ...Battery Free Portable Dental X ray USER MANUAL Doc UM 02FC 01 Rev 5 2021 01 29...
Страница 8: ...page 8 62 1 4 2 Labels and label positions Device labels 55x80mm NR F350 and CS 2400P...
Страница 10: ...page 10 62 Adaptor Label 38 X 90mm Warning Label 50 X 20mm...
Страница 11: ...page 11 62 Caution Label 60 X 20mm...