5.3 Maintenance
NANORAY
requires periodic constancy tests to ensure image quality and the safety
for the patient and operator.
Only
NANORAY
authorized technicians can perform inspection and service of this
equipment. For the technical assistance, contact
NANORAY
service center or your local
NANORAY
representative.
Our products are fast-charging products using Super Capacitor without using a typical
lithium-ion battery, so if they are not used within 2 hours after full charging, they must be
charged.
Note:
If the product has a failure or malfunction that requires repair, it must be
repaired by the manufacturer
Cautions and Notes
1)
Maintenance Task Checklist
Tasks
Period
Before operation, ensure that the equipment is clean and ready for use.
Daily
After using the equipment, make sure that the equipment has been turned
off.
Daily
Wipe the outer covers of the equipment with a dry cloth at the end of each
day’s operation.
DO NOT use detergents or solvents to clean the outer covers
of the equipment.
Daily
DO NOT keep the equipment or its parts in a humid place or near a
liquid substance.
Avoid placing the equipment near chemical storage and gas-filled
storage facilities.
Always turn off the equipment before performing any maintenance.
Tasks
Period
Ensure that the signal is audible and the X-ray emission light is visible
when you make an exposure.
Daily
Ensure that the yellow (exposure) indicator light turns on when the
Exposure Button is pressed.
Daily
Ensure that the battery charging LED indicator comes on when charging
the battery.
Daily
Ensure that all visible labels are intact and legible.
Monthly
Содержание Nanoray CS 2400P
Страница 1: ...Battery Free Portable Dental X ray USER MANUAL Doc UM 02FC 01 Rev 5 2021 01 29...
Страница 8: ...page 8 62 1 4 2 Labels and label positions Device labels 55x80mm NR F350 and CS 2400P...
Страница 10: ...page 10 62 Adaptor Label 38 X 90mm Warning Label 50 X 20mm...
Страница 11: ...page 11 62 Caution Label 60 X 20mm...