8400-M022-O1 Rev K
Model 190 Indicator
106
Cont2=
(Continuous Output on Option Card)
Press the
TARE
key to show the current setting. If the setting displayed is
no
(no),
press the
TARE
key again to save it.
Otherwise, use the
Fn/
key to toggle the setting to
no
(No Continuous Output on Option
Card) and press the
TARE
key to save it.
NOTE:
The default continuous output is
no
(No Continuous Output on Option Card) and
should not
be changed.
Ethernet Cable Installation
Connections to the 190 are by CAT 5 or equivalent cable. Consult your network
administrator for the proper procedure to terminate the cable.
After the cable has been terminated, plug the cable into the Ethernet port on the 190-IP.
IP Address Setup
Before operation may begin the Ethernet Device Server must have an IP Address. The
190-IP option card IP address is set using the DeviceInstaller by Lantronix, available when
logged into the dealer-protected software section of CardinalScale.com. Download the
application and its user’s guide from CardinalScale.com by navigating to 190 Storm >
Resource Type > Software > Lantronix DeviceInstaller for Cardinal IP Devices.
190-IP Operation
A client TCP connection to the 190-IP IP address at its listening port, 10001, can be used to
send commands to the 190. A description of the available commands is described in the
ASCII Commands section.