8400-M022-O1 Rev K
Model 190 Indicator
31
Calibration “
C
” Numbers
IMPORTANT! If any components have been changed that affect calibration
and/or your scale is used in a commercial application and must be "Legal for
Trade" you cannot use the “
C
” numbers to re-calibrate.
The “
C
” numbers are displayed only during the Test mode operation by pressing the
Fn/
key then the
UNITS/
key. The "
C
" numbers are shown at the end of the test operation
and each number is displayed for approximately 4 seconds, allowing you to record them.
Each number may be up to three (3) digits in length. By recording these numbers, you will be
able to return the indicator to its present calibration settings without using test weights simply
by entering the “
C
” numbers.
1.
With
CAL1=
displayed, press
UNITS/
key.
2.
At
C1=
prompt, press
TARE
key to show the current value of
C1
number.
3.
If the
C1=
value displayed is acceptable, press
TARE
key again to save it.
4.
Otherwise, use
Fn/
and
UNITS/
keys to enter a new
C1=
value and press the
TARE
key.
5.
At
C2=
prompt, press
TARE
key to show the current value of
C2
number.
6.
If the
C2=
value displayed is acceptable, press
TARE
key again to save it.
7.
Otherwise, use
Fn/
and
UNITS/
keys to enter a new
C2=
value and press the
TARE
key.
8.
At
C3=
prompt, press
TARE
key to show the current value of the
C3
number.
9.
If the
C3=
value displayed is acceptable, press
TARE
key again to save it.
10.
Otherwise, use
Fn/
and
UNITS/
keys to enter a new
C3=
value and press
TARE
key.
11.
At
C4=
prompt, press the
TARE
key to show the current value of the
C4
number.
12.
If the
C4=
value displayed is acceptable, press the
TARE
key again to save it.
13.
Otherwise, use
Fn/
and
UNITS/
keys to enter a new
C4=
value and press
TARE
key.
14.
The display will change to show:
SetgC?
.