8400-M022-O1 Rev K
Model 190 Indicator
76
Error Codes
The Cardinal 190 Weight Indicator is equipped with software that indicates when an error in
the operation takes place. The following lists the error codes displayed by the 190 along with
their meaning. Should you encounter an error code, please refer to this list for the cause.
Error
Cause (A.) and Corrective Action (B.)
ConFig
A.
Calibration required.
B.
Calibrate the scale with test weights.
ErrA h
A.
The analog input is off the scale at the upper end of the input range.
B.
Check wiring and load cell.
ErrA L
A.
The analog input is off the scale at the lower end of the input range.
B.
Check wiring and load cell.
Error
A.
Attempt to perform an illegal function:
1.
Performing a hold operation when the weight is invalid (e.g.,
overcapacity).
2.
Attempting to clear Tare with Gross weight greater than zero (
lft=1
,
usa=0
).
3.
Attempting to zero the scale when there is motion present or the scale is
over capacity.
4.
Attempting to zero the scale when Gross weight is out of zero range
(
trL=yES
).
5.
Attempting to print when the scale is over capacity.
6.
Attempting to do a span adjustment with insufficient load on the scale.
7.
Attempting to tare scale when there is motion present or the Gross
weight is below zero or over capacity.
Lo bAt
A.
This message is used with the battery operation and will turn ON to indicate
the battery does not have a sufficient charge and requires recharging.
If continued use furthers drains the battery, no change in operation will
occur until just before the battery voltage drops to a level where operation is
affected. At this level, the indicator will automatically turn off.
B.
Charge the battery pack or replace it.
notArE
The
NET/GROSS
key was pressed when there is no stored tare.
oCAP
A.
The load on the scale exceeds the capacity of the scale.
B.
Remove the excess load.
-oF-
A.
The indicator is trying to display a positive number with more than six digits
or a negative number with more than five digits.
B.
Return to Gross weight mode and review the Tare weight. Might indicate
incorrect calibration.
UnStb
A.
Motion was present when a print was initiated.
B.
Wait for a stable indication before printing.