8 | Installation
5.5 cu.ft. Undercounter Refrigerator
Installation
Leveling
Adjust the leveling feet located underneath the refrigerator by
turning clockwise to extend the feet, counter clockwise to
shorten them. It is important that the unit is leveled.
Intended Use
Do not locate or store your refrigerator outdoors. This model
was designed for indoor use only and violation of this will void
the terms of your warranty.
The refrigerators described in this manual are high
performance units for professional use. These products are
intended for use as cold storage in research use and as a
general purpose laboratory refrigerator, storing samples or
inventory at operating temperatures between 3°C and 7°C.
It is not considered a medical device and has therefore not
been registered with a medical device regulatory body
(e.g. FDA): that is, it has not been evaluated for the storage of
samples for diagnostic use or for samples to be re-introduced
into the body.
This unit is not intended for use in classified hazardous
locations, nor to be used for the storage of flammable or
corrosive inventory.
Providing Ventilation
Your Premier Plus undercounter refrigerator is equipped with
a ventilation area just above the door handle. Do not block or
cover the venting area as this can impact the performance of
the refrigerator.
Power On and Power Cord
Guidelines
1. Use only the power cord provided with your Premier Plus
undercounter refrigerator.
2. Plug the power cord into the receptacle on the upper rear
of the refrigerator as indicated. Secure the strain relief to
ensure the cord cannot accidentally pull out (as shown
above).
3. Position the refrigerator where you intend to locate it,
taking care to give it the recommended clearance and
plug into a grounded three prong receptacle. This outlet
must be within reach to disconnect power source.
The fans on your product will run at intervals of
variable speeds and may turn off completely.
They do not run constantly. If the fans run
frequently, the product may require more space
for ventilation. Please refer to the
“
” section on how to clean and
troubleshoot the fan for more information.
WARNING:
Fan has moving parts, product has
uninsulated live parts and hot parts. Do not
remove the top panel or attempt repairs. For
repairs or operating issues call Customer Care at
866-984-3766 for guidance and instructions.