background image

 

 
 
 

 

 

 

PHARMACY REFRIGERATOR 

 

 

Model No.:  ACR1515SS 

 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

Instruction Manual 

 

 

BEFORE USE, PLEASE READ AND FOLLOW ALL SAFETY RULES AND OPERATING INSTRUCTIONS. 

 
 
 
 

FELIX STORCH, INC. 
Summit Appliance Division 
770 Garrison Avenue 
Bronx, NY 10474 
www.summitappliance.com 

Write Serial No. (on back of unit) here: 

 
________________________________ 

ACCUCOLD™ is a division of Summit Appliance 

Содержание ACR1515SS

Страница 1: ...SE READ AND FOLLOW ALL SAFETY RULES AND OPERATING INSTRUCTIONS FELIX STORCH INC Summit Appliance Division 770 Garrison Avenue Bronx NY 10474 www summitappliance com Write Serial No on back of unit here ________________________________ ACCUCOLD is a division of Summit Appliance ...

Страница 2: ...llation Instructions 7 9 Location 7 Installing Shelves 8 Electrical System 8 9 Refrigeration Lighting System 9 10 Temperature Control 10 Lighting 10 User Maintenance 10 11 Cabinet Cleaning 10 Condenser Cleaning 10 Cleaning the Compressor Fan Compartment 11 Defrosting 11 Security Lock 11 Energy Saving Tips 11 Troubleshooting 12 Limited Warranty 13 14 2 ...

Страница 3: ...uctions are not followed IMPORTANT SAFEGUARDS Before the appliance is used it must be properly positioned and installed as described in this manual so read the manual carefully To reduce the risk of fire electrical shock or injury when using the appliance follow basic precautions including the following Plug into a grounded 3 prong outlet do not remove grounding prong do not use an adapter and do ...

Страница 4: ...d in accordance with the National Electrical Code and local codes and ordinances is required Do not kink or pinch the power supply cord of the appliance The size of the fuse or circuit breaker should be 15 amperes It is important that the appliance be leveled in order to work properly You may need to make several adjustments to level it All installations must be in accordance with local plumbing c...

Страница 5: ...EP HANDS AWAY from door pivot when door is open Before loading refrigerator CHECK that shelves are fixed and have the number of supports indicated for your model in the Instruction Manual CHECK that the front louver panel and rear grating are in place before plugging in unit DO NOT place chemical substances solvents reagents or food in this refrigerator This equipment is intended for the storage a...

Страница 6: ...Ventilation louvers Thermostat control Compressor and fan behind louvers Leveling legs HANDLING TRANSPORT DO NOT MOVE UNIT WITH PRODUCT INSIDE as loose product can break spill damage the appliance etc WHEN HANDLING STATIC CONDENSER MODELS SUCH AS THIS MAKE SURE THAT DOOR PIVOT IS LOCATED AS SHOWN IN FIGURE ABOVE 6 DO NOT TIP UNIT AT MORE THAN A 45 ANGLE DO NOT EXPOSE TO SUNLIGHT OR STORE AT TEMPER...

Страница 7: ...e free standing and not built into an enclosure Do not place the unit near heat sources such as stoves grills radiators etc and do not place in areas where direct sunlight may reach the unit CAUTION This appliance is designed for storing and cooling pharma ceuticals vaccines and other medical specimens Do not use it to store perishable food or chemical substances like solvents and reagents LOCATIO...

Страница 8: ...ORTANT TO AVOID ACCIDENTS WHEN CLEANING INSIDE OF UNIT FIRST REMOVE SHELVES AND SHELF CLIPS ELECTRICAL SYSTEM Proper electrical installation is required for your refrigerator to function correctly and to prolong the useful life of electrical parts and compressor BEFORE CONNECTING UNIT FOR FIRST TIME Verify that line voltage does not fluctuate more than 10 above the nominal value of 115 V 60 Hz DO ...

Страница 9: ...sures inside the refrigeration system to equalize and the oils in the compressor to settle After plugging the unit in let it run for one more hour without load or product The refrigerator may then be loaded DO NOT TOUCH THE COMPRESSOR TOP COVER while unit is running It is designed for high operational temperatures and its surface can become very hot which may cause burns The compressor is located ...

Страница 10: ...provide optimal energy consumption and serviceability It is recommended however that the unit be kept clean and that the following preventative maintenance be performed CABINET CLEANING Turn off the power unplug the appliance and remove all items including shelves and shelf clips When cleaning cabinet interior use a solution of warm water and mild soap Do not use steel wool caustic soap abrasive c...

Страница 11: ...t the drain hose is always inside the evaporation pan SECURITY LOCK Your unit is provided with a lock and key combination The lock may be found on the door frame just below the door handle The keys are located inside the plastic bag that contains the Instruction Manual After a key has been used remove it and put it in a secure place for safekeeping ENERGY SAVING TIPS DO NOT DISCONNECT UNIT AT NIGH...

Страница 12: ... does not seal properly Condenser needs cleaning The light does not work Not plugged in Tripped circuit breaker or blown fuse The light switch is OFF Ice forms on evaporator High relative humidity in environment The door is not closed properly The door is opened too often or for too long Interior air flow is blocked Vibrations Appliance is not level Fan blades are out of balance Tubes or moving pa...

Страница 13: ...ssors determined by SUMMIT to have been defective within this extended time period will at SUMMIT s option be either repaired or replaced with a compressor or compressor parts of similar design and capacity The above said extended compressor warranty applies only to hermetically sealed parts of the compressor and DOES NOT apply to any other parts or components including but not limited to cabinet ...

Страница 14: ...ranty specifically described above These warranties are exclusive and in lieu of all other warranties including implied warranty and merchantability or fitness for a particular purpose There are no other warranties which extend beyond the description on the face hereof OUTSIDE CONTIGUOUS UNITED STATES This warranty does not apply to and SUMMIT is not responsible for any warranty claims made on pro...