10 | Operation
5.5 cu.ft. Undercounter Refrigerator
Operation
LED Indicator
When you turn on your refrigerator, verify that the LED
indicator is working. The LED indicator appears on the front
control panel next to the upper corner of the display. The LED
indicator illuminates green when the unit is on and running. It
illuminates red if one of the following conditions are present:
•
Low Battery
•
Full Memory
•
Data Logging Error
Checking the Glycol
Sample Bottle
The glycol sample bottle is pre-installed inside your unit. The
sample bottle allows you to monitor the internal temperature
of items being stored in the refrigerator.
Verify the following:
1. The cord for the sample bottle is fully plugged into the
back wall of the interior of your unit.
2. The sample bottle is securely mounted in the clip
provided.
3. The level of the propylene glycol in the bottle is full.
In the event that the bottle has broken during shipping,
contact our service department for a replacement bottle.
Preprogrammed Settings
Settings can be customized via the Settings and Alerts
menus.
Temperature Display & Alarm – The display is preset to read
the glycol sensor but can be changed to read the air sensor.
Temperatures are detected and displayed in Celsius units to
one tenth of a degree. For example, 4.0°C. Upper and lower
alarm thresholds are driven by the sensor selected for the
display.
•
Set Point Temperature – The unit preset/default is set at
5.0°C (39.2°F)
•
Upper Threshold Alarm – The unit preset/default is set at
8.0°C (46.4°F)
•
Lower Threshold Alarm – The unit preset/default is set at
2.0°C (35.6°F)
Preset Alarms/Alerts
The following alarms/alerts are factory preset:
A power outage would result in a flashing red or
green light every 15 seconds. Flashing green
indicates continuous data logging. Flashing red
indicates a data logging error. See the
troubleshooting guide for ways to resolve the
issue.
Alarm
Factory Setting
Icon
Temperature Range
Alarm
min 2°C - max 8°C
Door Ajar Alarm
1 Minute
No Power Alarm
>10 seconds outage
Battery Low Alarm
<30% remaining
Memory Full Alarm
<30% remaining