CHAPTER 5 TROUBLESHOOTING
Description of the Scaling Control Screen
Figure 5-228
No. Details
{
1
Factory Scale Parameter
The factory default scale parameter
setting value.
Can be changed using the slide bar.
Units are percentage of the entire length.
The setting range is ±1.0.
The [+] direction increases the length of
the image.
{
2
User Scale Parameter
This setting value is added to
{
1
above.
The setting range is ±3.0.
Other details are the same as
{
1
.
The setting value of
{
1
+
{
2
is applied.
{
3
Registration
On: The setting value of the reading
start position.
Off: The setting value of the reading end
position.
Can be changed by directly entering in
the data box or by using the scroll
arrows.
Units are mm, and the setting range is
±5.0.
The [+] direction delays the timing.
For example, if set to On[1.0], 1mm of
the leading edge of the image is cut off.
Note:
In this machine, only “On” is used for
adjustment. And more than 2 mm in
the minus direction for “On” is invalid.
No.
Details
{
4
Set / Reset
Set: Sets the changed values.
Reset: Returns the setting values to
“0”.
{
5
Registration Adjustment
Click this button to execute
[Registration Adjustment].
{
6
Scanned Image
Displays the scanned image.
The left side is the front image and the
right side is the back image.
You can move the image using the
scroll bars.
{
7
Zoom
Enlarges the image in
{
6
above using
a slide bar.
{
8
Resolution / Size
Selects the resolution and size of the
scan using pull-down menus.
{
9
Scan
Click this button to begin the scan.
Table 5-203
{
1
{
2
{
3
{
4
{
5
{
6
{
7
{
8
{
9