Operating the Video Display
25
Performing calibrations (
You can perform calibration using an external sensor, without using a computer.
The supported external sensors are Konika Minolta Display Color Analyzers CA-310 and CA-210. Be sure to also read
the instruction manual of the CA-310 and CA-210.
1 Connect the display color analyzer to the USB port of the main unit.
2 Select [User 1] to [User 7] and then open the [Calibration] menu.
1
[Adjustment] [Picture Mode] [User 1–7]
2
[Adjustment] [Detail Settings] [Calibration]
•
Set each target value.
3 Press the jog dial and select [Start].
•
Please follow the information indicated on the screen.
4 Initialize the sensor.
•
Set the mode dial of the Universal Measuring Probe to [0-CAL].
•
Press the jog dial of the video display, select [OK], and execute initialization.
5 Place the universal measuring probe pointing at the center of the video display.
•
Set the mode dial of the universal measuring probe to [MEAS] and place the probe as shown below according to
the displayed content. Press the jog dial of the video display, select [OK], and execute calibration.
Konica Minolta Universal Measuring Probe
CA-310 support: CA-PU32, CA-PU35
CA-210 support: CA-PU12, CA-PU15
Konica Minolta
Display Color Analyzer CA-310, CA-210
Main Unit
6 Finish calibration.
•
When the message [Calibration is completed.] is displayed, press the jog dial and select [OK].
•
If the message [Calibration error.] is displayed.
Calibration has been terminated due to an error. The main unit returns to the state before calibration. (
•
To cancel calibration
Press the jog dial during calibration and select [Cancel]. The main unit returns to the state before calibration.