3-6) Create the graph below based on the data measured in step 5).
0.00
0.50
1.00
1.50
2.00
2.50
3.00
3.50
1
2
3
4
5
6
7
8
9
10
11
Step
Density
Max setting
density
Film base
density
As shown in the above graph, the measurement values need only to nearly form a straight line
from the minimum density to the maximum density.
The important point here is that the measurement values make a straight line and a maximum
density corresponding to the settings is output.
If the measurement values deviate too far from the straight line and a maximum density
corresponding to the settings is not output, printer and monitor output linearity settings,
calibration, and other adjustments are necessary.
* The above graph is an example of measurements when the maximum density is set to 3.2 D
and the printed film is measured with a densitometer. Refer to the data below when the
maximum density is set to 3.1 D, or when gradation analysis software is used to take
measurements on the monitor.
* The LINEAR output cannot be set for some printer models.
In this case, try to select an output setting as close as possible to LINEAR.
Ideal density values in LINEAR LINE
Step
1
2
3
4
5
6
7
8
9
10
11
3.2 D
0
0.32
0.64
0.96
1.28
1.60
1.92
2.24
2.56
2.88
3.20
3.1 D
0
0.31
0.62
0.93
1.24
1.55
1.86
2.17
2.48
2.79
3.10
12 Bit
0
410
819
1229
1638
2048
2457
2867
3276
3686
4095
8 Bit
0
26
51
77
102
128
153
179
204
230
255
Possible causes of non-linearity in measurement values:
* Inadequate calibration of printer and high-definition monitor
* Inadequate settings for CXDI printer parameters
* Faults in printer or high-definition monitor
Содержание CXDI-50G
Страница 8: ...1 General 2 2 2 S Sy ys st te em m D Di ia ag gr ra am m 2 1 Standalone System 2 2 Total System ...
Страница 20: ...2 Installation Manual 3 CXDI 50G Power Box assemble package 4 Grid optional 6 ...
Страница 32: ...2 Installation Manual When normal imaging When timeout due to RX_REQ not negating 18 ...
Страница 58: ...2 Installation Manual 44 Input parameters Fig 3 ...
Страница 104: ...2 Installation Manual 90 2 Power Box Unit mm Mass 4 2Kg ...
Страница 105: ...2 Installation Manual 91 3 Remote Switch Unit mm Mass 0 5Kg 4 Grid Optional 508 27 452 Unit mm Mass 0 9Kg Frame ...
Страница 135: ...CXDI SYSTEM CXDI 50G 1 2 Rev 02 Key No 5 Key No 6 Key No 7 Key No 8 Key No 9 Key No 10 Key No 11 Canon Inc ...
Страница 139: ...CXDI 50G IMAGING UNIT 2 4 Rev 02 16 21 19 18 22 20 17 2 4 3 8 10 12 13 9 11 15 1 5 6 14 7 Canon Inc ...
Страница 150: ...Content 1 Gathering Information about Problems 1 2 How to back up and recover the system 3 CXDI V6 00 Error code ...