![Canon CXDI-50G Скачать руководство пользователя страница 176](http://html.mh-extra.com/html/canon/cxdi-50g/cxdi-50g_service-manual_61474176.webp)
ror Code List for Versi
Error Code List for Version 6.00
Error Code
Seriousness
Reply
Keys
Message
[Cause]
[Operation after Reply Key is touched]
-517
Warning
RETRY #1:Warning
#2:Internal Storage Error:
#3:Processed image cannot be stored, because the capacity of
internal storage memory is insufficient.
[Cause]
The space in the internal disk is insufficient.
[Operation after Reply Key is touched]
QA process will be performed again.
-518
Warning
RETRY
CANCEL
#1:Warning
#2:Internal Storage Error:
#3:DICOM image cannot be stored, because the capacity of
internal storage memory is insufficient.
[Cause]
There is not enough space for storing DICOM image.
[Operation after Reply Key is touched]
When RETRY is chosen, image will be stored into the
internal disk.
When CANCEL is chosen, storing of image will be cancelled
after displaying a confirmation message.
-519
Warning
RETRY
CANCEL
#1:Warning
#2:External Storage Error:
#3:Capacity of external storage memory is insufficient. Retry
after making a space in the disk.
[Cause]
There is not enough space in the external storage medium.
[Operation after Reply Key is touched]
When RETRY is chosen, image will be stored into the
external storage medium.
When CANCEL is chosen, storing of image will be cancelled
after displaying a confirmation message.
-520
Warning
OK #1:Warning
#2:Internal Storage Error:
#3:Exposed image cannot be stored, because the capacity of
internal storage memory is insufficient.
[Cause]
Space in the internal disk is insufficient.
[Operation after Reply Key is touched]
When RETRY is chosen, the image will be stored into the
internal disk.
When CANCEL is chosen, storing of image will be cancelled
after displaying a confirmation message.
- 21 -
Содержание CXDI-50G
Страница 8: ...1 General 2 2 2 S Sy ys st te em m D Di ia ag gr ra am m 2 1 Standalone System 2 2 Total System ...
Страница 20: ...2 Installation Manual 3 CXDI 50G Power Box assemble package 4 Grid optional 6 ...
Страница 32: ...2 Installation Manual When normal imaging When timeout due to RX_REQ not negating 18 ...
Страница 58: ...2 Installation Manual 44 Input parameters Fig 3 ...
Страница 104: ...2 Installation Manual 90 2 Power Box Unit mm Mass 4 2Kg ...
Страница 105: ...2 Installation Manual 91 3 Remote Switch Unit mm Mass 0 5Kg 4 Grid Optional 508 27 452 Unit mm Mass 0 9Kg Frame ...
Страница 135: ...CXDI SYSTEM CXDI 50G 1 2 Rev 02 Key No 5 Key No 6 Key No 7 Key No 8 Key No 9 Key No 10 Key No 11 Canon Inc ...
Страница 139: ...CXDI 50G IMAGING UNIT 2 4 Rev 02 16 21 19 18 22 20 17 2 4 3 8 10 12 13 9 11 15 1 5 6 14 7 Canon Inc ...
Страница 150: ...Content 1 Gathering Information about Problems 1 2 How to back up and recover the system 3 CXDI V6 00 Error code ...