ror Code List for Versi
Error Code List for Version 6.00
Error Code
Seriousness
Reply
Keys
Message
[Cause]
[Operation after Reply Key is touched]
-521
Error
OK #1:Error
#2:Queue File Open Error:
#3:Queue file is not found.
[Cause]
Failed to open the queue file. This error occurs when no
queue file exists.
[Operation after Reply Key is touched]
*Under the current system, this error is unlikely to occur.
-599
Error
OK #1:Error
#2:File Open Error:
#3:File is not found.
[Cause]
This error occurs when the system has failed to open the file
during RESEND.
[Operation after Reply Key is touched]
Just touch OK.
The system returns to the state before the occurrence of the
error.
-601
Error
OK #1:Error
#2:Resend Error
#3:File may have been automatically deleted.
[Cause]
The image cannot be resend as it was deleted for some reason
while being called out.
[Operation after Reply Key is touched]
The image being displayed on the re-print screen is made
invalid and the image can no longer be resend.
-5100
Error
OK #1:Error
#2:Sensor Unit:Detect Error
#3:Check the sensor power unit or cable connection. If you
want to recover the system, shut it down.
[Cause]
The power of the imaging unit is OFF, or the cable is
disconnected or damaged.
[Operation after Reply Key is touched]
The system only requires your confirmation. Touch OK. Only
the problematic sensor is disabled.
- 14 -
Содержание CXDI-50G
Страница 8: ...1 General 2 2 2 S Sy ys st te em m D Di ia ag gr ra am m 2 1 Standalone System 2 2 Total System ...
Страница 20: ...2 Installation Manual 3 CXDI 50G Power Box assemble package 4 Grid optional 6 ...
Страница 32: ...2 Installation Manual When normal imaging When timeout due to RX_REQ not negating 18 ...
Страница 58: ...2 Installation Manual 44 Input parameters Fig 3 ...
Страница 104: ...2 Installation Manual 90 2 Power Box Unit mm Mass 4 2Kg ...
Страница 105: ...2 Installation Manual 91 3 Remote Switch Unit mm Mass 0 5Kg 4 Grid Optional 508 27 452 Unit mm Mass 0 9Kg Frame ...
Страница 135: ...CXDI SYSTEM CXDI 50G 1 2 Rev 02 Key No 5 Key No 6 Key No 7 Key No 8 Key No 9 Key No 10 Key No 11 Canon Inc ...
Страница 139: ...CXDI 50G IMAGING UNIT 2 4 Rev 02 16 21 19 18 22 20 17 2 4 3 8 10 12 13 9 11 15 1 5 6 14 7 Canon Inc ...
Страница 150: ...Content 1 Gathering Information about Problems 1 2 How to back up and recover the system 3 CXDI V6 00 Error code ...