2. Installation Manual
- 47 -
Parameter List (Separate Document 1)
DICOM storage device
In the CXDI, DICOM data transfer is performed using the transfer software module
“send_image” .The settings for these parameters are described below.
Parameter Meaning
Description
-m maxPDU
Maximum PDU value in
byte units
* The CXDI automatically
uses 131072 internally for
operation.
* Designating a specific value
allows overwriting of the
above value.
* The DICOM standards do not allow
values of 1301073 or higher to be set.
* This is used when the operator who
manages the connected storage device
requests a size change.
* In DICOM printing, note that the
argument title changes to -u.
(
Æ
See the printing parameters.)
-t calledTitle
Called App Entity Title
* The AE Title setting field is
automatically applied to this
setting.
* Designating a specific value
allows overwriting of the
above value.
* Note that the meaning is opposite of the
DICOM printing argument -c.
(
Æ
See the printing parameters.)
* This is used when the operator who
manages the connected storage device
requests a change in the installed
identification information (version).
-c callingtitle
calling App Entity Title
* The CXDI automatically
uses CANON_CCR
internally for the operation.
* Designating a specific value
allows overwriting of the
above value.
* Note that the meaning is opposite of the
DICOM printing argument -c.
(
Æ
See the printing parameters.)
* This is used when the operator who
manages the connected storage device
requests a change in the installed
identification information (version).
-s SOPName
(for reference)
This parameter designates
whether class be connected
for performing association
at the beginning of
transfer.(CR/T/MR/NM/S
C/US)
* This is not used in the
CXDI.
-I
A-RELEASE-RES is
ignored.
* This parameter is used
simply as “-I”
* This is used when the error message
“30012 Peer aborted Association (or
never connected)” occurs even though
the DICOM data transfer was
successful.
Æ
This is used differently based on the
connected storage devices.
-d FAC
This parameter dumps a
specific facility log.
(DCM/DUL/SRV)
* This parameter is used
simply as “-d”
* This parameter is used to
make the transfer software
put the debugging character
string on the console.
* This parameter does not affect DICOM
data transfer.
* This parameter outputs the CXDI log
based on Windows NT.
Содержание CXDI-50G
Страница 8: ...1 General 2 2 2 S Sy ys st te em m D Di ia ag gr ra am m 2 1 Standalone System 2 2 Total System ...
Страница 20: ...2 Installation Manual 3 CXDI 50G Power Box assemble package 4 Grid optional 6 ...
Страница 32: ...2 Installation Manual When normal imaging When timeout due to RX_REQ not negating 18 ...
Страница 58: ...2 Installation Manual 44 Input parameters Fig 3 ...
Страница 104: ...2 Installation Manual 90 2 Power Box Unit mm Mass 4 2Kg ...
Страница 105: ...2 Installation Manual 91 3 Remote Switch Unit mm Mass 0 5Kg 4 Grid Optional 508 27 452 Unit mm Mass 0 9Kg Frame ...
Страница 135: ...CXDI SYSTEM CXDI 50G 1 2 Rev 02 Key No 5 Key No 6 Key No 7 Key No 8 Key No 9 Key No 10 Key No 11 Canon Inc ...
Страница 139: ...CXDI 50G IMAGING UNIT 2 4 Rev 02 16 21 19 18 22 20 17 2 4 3 8 10 12 13 9 11 15 1 5 6 14 7 Canon Inc ...
Страница 150: ...Content 1 Gathering Information about Problems 1 2 How to back up and recover the system 3 CXDI V6 00 Error code ...