10
CT-500 Instruction Manual
1.
Place the bath control housing on the surface where you plan to operate the instrument. This
should be a sturdy, level tabletop or bench with a non-flammable surface.
Caution:
Use caution when lifting the bath control housing as it is quite heavy.
2.
Place the felt pad(s) on top of the insulating disk. The CT-600 requires two felt pads, the CT-500
requires one felt pad.
3.
Place the glass jar on top of the felt pad(s).
Caution:
Use caution when lifting the bath jar as it is quite heavy.
4.
Position the rear cover containing the motor housing, heaters, cooling coil, and baffle on top of
the jar, feeding all connecting cables from the heaters, float switch and control/overtemp
probes through the vertical wiring channel in the rear of the housing. The back of the motor
housing should fit over the vertical channel, and the groove on the underside of the half-round
cover should fit over the edge of the jar. Refer to Figure 3.
Figure 3: Position the Rear Cover
5.
Place the front bath cover (with holes for the viscometers) on top of the front half of the jar.
Refer to Figure 4.