CANNON Instrument Company | CT-500 Operation
23
To heat the bath more rapidly, turn on the
PREHEAT
(
PH
) switch. If the target temperature is below
80 °C, turn the
PH
switch off when the desired temperature is reached. For temperatures above
80 °C, keep the
PH
switch ON to maintain temperature control. When the
PH
switch has been
turned on, the
PH - HEAT
LED will light.
3.
As the bath temperature approaches the selected
T - SELECT
setting, the LED above the
TA
controls will start to blink as the heater duty cycle is automatically adjusted to stabilize
temperature.
If the auxiliary heater was engaged, the
PREHEAT
light will also blink. Make sure to turn off the
PREHEAT
switch if operating at temperatures below 80 °C.
4.
Insert the appropriate thermometer for the test temperature (see
5.
To adjust the bath temperature to the exact temperature desired, first unlock the
TA - FIXED
dial.
6.
Read the thermometer to determine the actual temperature of the bath. Make sure that both
the bath temperature and thermometer have stabilized before taking a reading.
7.
Turn the
TA - FIXED
dial to adjust the temperature. After each adjustment, allow several minutes
for the bath temperature to stabilize.
8.
Turn the dial clockwise
to increase the bath temperature or turn the dial counter-clockwise
to
decrease the bath temperature.
9.
When the desired temperature has been reached, re-lock the
TA - FIXED
dial.
10.
After the bath has reached the desired temperature, set the
HIGH TEMPERATURE LIMIT
CONTROL (LC)
. This will prevent the bath from overheating if a malfunction occurs.
Adjusting the High Temperature Limit Control
During this procedure, the temperature of the bath may change slightly, but will quickly recover.
1.
Using a screwdriver, slowly turn the
LC - TEMPERATURE ADJUST
control counterclockwise until
the OVER TEMP message lights up on the
LC - PUSH TO RESET
button.
2.
Turn the control clockwise approximately ¼ to ½ turn.
3.
Push the
LC - PUSH TO RESET
button. If the bath doesn't recover, repeat step 2 and try again.
Содержание CT-500
Страница 1: ...25 9991 Instruction Operation Manual Constant Temperature Bath CT 500 600 Series ...
Страница 2: ......
Страница 4: ......
Страница 8: ......