Maintaining the System

 

• Only authorized Syneron-Candela technicians are qualified to service the system.

 

• Shut down the system and disconnect the power cable before performing any mainte   

 

nance procedures (see Chapter 7).

Fire Hazards

 

• The absorption of optical energy or conduction of RF raises the temperature of the  

 

 

absorbing material. Do not use the system in the presence of explosive or flammable  

 

 materials.

 

• Keep drapes and towels moist to prevent them from igniting and burning. Use non-flam  

 

mable prepping solutions.

 

• Do not use flammable substances when preparing the skin for treatment.

 

• If alcohol is used for disinfecting, it must be allowed to dry thoroughly before the system  

 

is used.

System Classifications

 

• Electric shock protection: Class I, Type BF

 

• Protection against ingress of liquids: Ordinary equipment

 

• Not suitable for use in presence of flammable substances

Warning

No modifications to this equipment are allowed!

Содержание VelaShape

Страница 1: ...VelaShape Copyright 2020 Manufacturer s User Instruction Manual...

Страница 2: ...Copyright 2020 Product Guide All Rights Reseved...

Страница 3: ...Copyright 2020 Product Guide All Rights Reseved...

Страница 4: ...hape II system must be serviced only by Syneron Candela qualified personnel Conventions Used in this Manual The following conventions in the form of notes cautions and warnings are used in this manual...

Страница 5: ...Copyright 2020 Product Guide All Rights Reseved Explanation of the Symbols Used in the System...

Страница 6: ...provided in this chapter Global Wellness Enterprises assumes no liability what so ever for any damage or injury as a result of an application of a product which is not in strict accordance with the i...

Страница 7: ...attendant and the patient Follow these personal safety guidelines to maintain a high level of personnel safety Make sure that only well trained attendants perform treatment Prior to scheduling treatme...

Страница 8: ...of explosive or flammable materials Make sure that all treatment personnel are familiar with the system s controls and know how to shut down the system instantly Infrared Energy Safety Issues The Vela...

Страница 9: ...hair etc in the treatment area Ensure integrity of rollers coating before and during treatment Electrical and Mechanical Safety Keep all covers and panels of the system closed Removing the covers cre...

Страница 10: ...ot use the system in the presence of explosive or flammable materials Keep drapes and towels moist to prevent them from igniting and burning Use non flam mable prepping solutions Do not use flammable...

Страница 11: ...Guide All Rights Reseved System Nameplate The following illustration see Figure 1 displays the location of the system s nameplate which includes the system s identification regulatory and electrical r...

Страница 12: ...for the relief of minor muscle aches and spasms as well as for the temporary improvement of local blood circulation The system is designed to be user friendly safe and reliable It integrates the foll...

Страница 13: ...Please note that working simultaneously with all 3 applicators is not possible The VelaShape applicators are composed of a permanent handpiece and a replaceable cover The replaceable covers of the Vel...

Страница 14: ...tor panel of each applicator The three user defined treatment modes are IR RF and Vacuum The levels are indicated by 0 1 2 or 3 illuminated LEDs which correspond to no energy low medium and high level...

Страница 15: ...skin will result in disabled IR and RF energies indicated by Blinking of the LEDs on the applicator s operator panel An Improve Contact message on the communication terminal s display see Figure 5 Com...

Страница 16: ...c effect of the IR and RF responds in a higher diffusivity of oxygen This in turn increases the metabolic breakdown of stored fat by the fat cells lipolysis thereby reducing the size of the fat cells...

Страница 17: ...Guide All Rights Reseved Figure 8 Before Treatment Figure 9 IR Heats the Tissue Up To 2 mm RF Heats the Tissue Up To 20 mm in Depth Figure 10 After VelaShape Treatment Reduction in the Size of the Fat...

Страница 18: ...e RF energy reaches the depth of the hypodermis Mechanical Effects on the Skin Mechanical Rollers see Figure 9 Rollers knead the skin inward so that the IR and RF energies can penetrate deeper in the...

Страница 19: ...then continue with unpacking and installation from this point 4 Carefully remove the packed accessories and small Styrofoam boxes from inside the Styrofoam covers 5 Finish removing all of the Styrofoa...

Страница 20: ...bly Vcontour applicator cradle Umbilical support mast includes five harness loops Accessories kit including Communications terminal with cable User Manual Quick Reference Guide Quick Service Guide Inf...

Страница 21: ...n the following manner 1 Unpack the umbilical support mast with the five installed harness loops see Figure 12 Ensure that the harness loop s Allen screws are properly seated in their dimples in the m...

Страница 22: ...nsert the mast as shown in Figure 13 until it is stopped by the pin Figure 14 Vcontour Applicator Cradle Figure 13 Inserting the Umbilical Support Mast 4 Attach the Vcontour applicator cradle to the m...

Страница 23: ...connection port see Figure 15 Figure 16 Inserting the Dual Connector Assembly Figure 15 Dual Connector Assembly Electronic Vacuum Sockets 6 Firmly insert the dual connector assembly into the system s...

Страница 24: ...by pressing its two clamps onto the mast as shown in Figure 17 Figure 18 Installing the Communication Terminal 2 Figure 17 Installing the Communication Terminal 1 8 Route the terminal s communication...

Страница 25: ...ction port on the system s rear panel and plug the cable into the appropriate electrical outlet 17 Refer to Figure 11 and ensure that all system components are correctly assembled 18 Insert the key in...

Страница 26: ...handle at the back of the system and carefully lift it up onto the vehicle Spread padding on the wall of the transport vehicle to avoid scratching the system s covers and strap the system to the wall...

Страница 27: ...rotection against fire replace the fuse only with one of the same type and rating Proper grounding is essential for safe operation Do not replace any electric part of the system except fuses Only auth...

Страница 28: ...cles located on the aperture guide can damage it therefore dust particles should be kept to a minimum For optimal operation of the system maintain room temperature between 15 30 C 59 86 F and relative...

Страница 29: ...present inside the system Always be aware of the possible dangers and take proper safeguards as described in this manual Always turn the system off before connecting disconnecting an applicator or Con...

Страница 30: ...e displayed in one of the following manners If a fault situation is displayed refer to Chapter 8 Troubleshooting 1 RF failure RF LEDs column is blinking 100 RF Failure message on the communication ter...

Страница 31: ...modes The operator panel is located on the top of both applicators The operator panel consists of three treatment mode buttons and informational display LEDs Each button on the operator panel applies...

Страница 32: ...High 3 Communication Terminal Messages During system operation the terminal s display LCD will always show a split view of the applicators operating conditions and other relevant information such as R...

Страница 33: ...ous Maintenance or Error messages will be discussed in the troubleshooting guide later in this manual see Chapter 8 Initiating Treatment Now that you have defined your treatment settings you are ready...

Страница 34: ...obtain better results but is not essential However weight gain may have a negative effect on the results Indications The VelaShape system is indicated for non invasive treatment for body contouring vi...

Страница 35: ...e healing Treating over tattoo or permanent makeup Area with varicose veins As per the practitioner s discretion refrain from treating any condition which might make it unsafe for the patient Although...

Страница 36: ...the treatment with clean dry skin no perfume no lotion Hair may impair the movement of the applicator overheat the treatment area or be pulled out during treatment Removing hair before treatment is a...

Страница 37: ...energy delivery You should examine the patient s skin before treatment to determine the suitable type of treatment to apply to ensure effective treatment with a minimal risk of skin damage Selecting t...

Страница 38: ...after the treatment session may be inaccurate due to a transient edema The suggested measurement method is 1 The patient should stand in an upright position adjacent to a wall barefoot and with both...

Страница 39: ...ne buttock etc Divide the area into strips dimensions of each strip width of the applicator the distance between the electrodes and the length of 30 cm 12 inches Vcontour Applicator the area dimension...

Страница 40: ...e this endpoint is reached this heat must be retained for at least 5 10 minutes for treatment to be effective Treat the desired area in the following manners see page 46 1 Gliding slowly in Back Forth...

Страница 41: ...times may not be effective It is recommended to treat the front thigh and abdomen from a supine position and the back thigh buttocks and the back from a prone position Very large hips and inner thighs...

Страница 42: ...natomical area using the motions described above Step 2 Return to the same anatomical area after 10 minutes and repeat the procedure reaching the endpoint again While waiting you may treat another are...

Страница 43: ...ht edema Blinking of the IR and or RF LEDs on the operator panel indicates inadequate coupling which results in RF and IR being disabled CAUTION Release the trigger before you lift the applicator from...

Страница 44: ...he patient feels an intolerable stinging sensation Low IR levels with higher vacuum levels enable prolonged and deeper treatment e g circumference reduction Higher IR levels with lower vacuum levels f...

Страница 45: ...eatment overheating e g Jacuzzi sauna etc or risk of physical damage Treatment Conclusion Treatment should be concluded when the results are satisfactory to the patient or according to the operator s...

Страница 46: ...ower cable s connection port indicated WARNING DISCONNECT POWER SUPPLY CABLE BEFORE MAINTENANCE OR SERVICE PROCEDURES TO AVOID ELECTRIC SHOCK For continued protection against fire replace the fuse onl...

Страница 47: ...impaired Roller axis appears to wear out color turns from silver to copper Proceed as follows 1 Turn the system off and disconnect the replaceable applicator cover by pressing the two spring loaded b...

Страница 48: ...ccording to the schedule printed on the cover s box or when the communication terminal displays a 03 Replace Filters message see Figure 25 Never use a pair of filters for more than 10 accumulative hou...

Страница 49: ...pplicator s Filters Caution Make sure that the interior of the applicator is completely dry before inserting the filters Cleaning and Maintaining the Vcontour Applicator Cleaning Replacing the Applica...

Страница 50: ...alcohol until all residual VelaSpray Ease is removed Wipe and dry the cover with a soft cloth Reconnect the cover only when completely dry 3 Clean the Vcontour applicator s components at the end of ev...

Страница 51: ...n terminal displays a 03 Replace Filters message 1 Use the supplied filter extractor to pull the filter out of the cover as shown in Figure 27 2 Wash the filter and allow it to dry completely before r...

Страница 52: ...daily and cleaned according to the amount of dust and debris that accumulates To clean the system air filter see Figure 29 1 Turn the system off 2 Open the filter door by lifting up the lock 3 Releas...

Страница 53: ...ist of possible system errors that will be displayed on the communication terminal s display and operator panels of the applicators and how to correct them If the corrective actions listed in the tabl...

Страница 54: ......

Страница 55: ...60 Hz single phase 110 127 VAC 5 2A 50 60 Hz single phase 220 240 VAC 3 1A 50 60 Hz single phase Environmental Requirements Treatment Room Operating temperature 15 30 C 59 86 F Relative Humidity Up to...

Страница 56: ...Protection Against Ingress of Water IPX0 Ordinary Equipment Dimensions VelaShape II system W x D x H 36 x 55 x 80 cm 14 2 x 21 6 x 31 5 Applicator umbilical harness 2 8 m 9 2 feet Weight VelaShape II...

Страница 57: ...All Rights Reseved Electromagnetic Emissions The VelaShape II system is intended for use in the following electromagnetic environment The owner of the system and or the user must ensure that it is ope...

Страница 58: ...quality should be that of a typical commercial or hospital environment If the VelaShape II system is required to operate during power interruptions the system should be powered from an uninterruptible...

Страница 59: ...electromagnetic environment due to fixed RF transmitters an electromagnetic site survey should be considered If the measured field strength in the location in which the VelaShape II system is used exc...

Страница 60: ...tic environment in which radiated RF disturbances are controlled The owner of the system and or the user can help prevent electromagnetic interference by maintaining a minimum distance between portabl...