GentleYAG Laser System
Candela Corporation
8501-00-1766 Revision F
CONFIDENTIAL
Page 79
14 Perform laser treatment.
Note: In some cases, when treating large areas, a PURGE fault may appear as
a result of the accumulation of reflected heat. If this happens, allow delivery
system to cool before continuing treatment. Place laser in STANDBY state
and check the windows on the distance gauge and handpiece slider. Dirty
windows will reflect laser energy back into the handpiece which will heat it
up and cause PURGE faults. If possible, replace with a cool delivery system.
16.
Place the laser into
STANDBY after use. Document laser use.
CAUTION!
Failure to initiate a calibration after cleaning/replacing the window or delivery
system may result in the delivery of excessive laser energy.
Содержание GentleYAG
Страница 1: ...Operator s Manual 0123 8501 00 1766 Revision F June 2006 ...
Страница 4: ...GentleYAG Laser System Candela Corporation 8501 00 1766 Revision F CONFIDENTIAL Page 4 ...
Страница 48: ...GentleYAG Laser System Candela Corporation 8501 00 1766 Revision F CONFIDENTIAL Page 48 ...
Страница 125: ...GentleYAG Laser System Candela Corporation 8501 00 1766 Revision F CONFIDENTIAL Page 125 Labeling Symbols Section 7 ...
Страница 133: ...GentleYAG Laser System Candela Corporation 8501 00 1766 Revision F CONFIDENTIAL Page 133 Label 13 ...
Страница 137: ...GentleYAG Laser System Candela Corporation 8501 00 1766 Revision F CONFIDENTIAL Page 137 5 A B C 6 ...
Страница 144: ...GentleYAG Laser System Candela Corporation 8501 00 1766 Revision F CONFIDENTIAL Page 144 ...
Страница 145: ...GentleYAG Laser System Candela Corporation 8501 00 1766 Revision F CONFIDENTIAL Page 145 Accessory List Section 8 ...
Страница 150: ...GentleYAG Laser System Candela Corporation 8501 00 1766 Revision F CONFIDENTIAL Page 150 ...